Analytical verification of Combur10 Test® M strips and urine analyzer Cobas u411


Vrtarić, Alen; Nikolac, Nora; Miler, Marijana; Šimundić, Ana-Maria
Analytical verification of Combur10 Test® M strips and urine analyzer Cobas u411 // Biochemia Medica / Šimundić, Ana-Maria (ur.).
Zagreb: Hrvatsko društvo za medicinsku biokemiju i laboratorijsku medicinu (HDMBLM), 2015. str. S100-S101 (poster, domaća recenzija, sažetak, znanstveni)


CROSBI ID: 820369 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Analytical verification of Combur10 Test® M strips and urine analyzer Cobas u411

Autori
Vrtarić, Alen ; Nikolac, Nora ; Miler, Marijana ; Šimundić, Ana-Maria

Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni

Izvornik
Biochemia Medica / Šimundić, Ana-Maria - Zagreb : Hrvatsko društvo za medicinsku biokemiju i laboratorijsku medicinu (HDMBLM), 2015, S100-S101

Skup
8. kongres Hrvatskog društva za medicinsku biokemiju i laboratorijsku medicinu s međunarodnim sudjelovanjem

Mjesto i datum
Rijeka, Hrvatska, 22.09.2015. - 26.09.2015

Vrsta sudjelovanja
Poster

Vrsta recenzije
Domaća recenzija

Ključne riječi
method verification; method comparability; precision

Sažetak
Introduction: Our aim was to: a) verify urinary test strips for urine chemistry analysis Combur10Test® M (Roche Diagnostics GmbH, Mannheim) on analyzer Cobas u411 ; b) asses the comparability of Cobas u411 and Miditron Junior II (both Roche Diagnostics, GmbH, Mannheim). Materials and Methods: Analytical verification was performed according to CLSI guideline EP12- A. Repetability was tested on 10 consecutively repeated measurements on negative control (BIORAD, Liquicheck Urinalysis Control Level 1) and in two patient samples with pathological values. Precision from day-to-day was assesed on control samples (Level 1 and 2) in duplicate for 10 days. Comparability was evaluated comparing 87 patient›s samples on analyzers Cobas U411 and Miditron Junior II. Accuracy for glucose, bilirubin and proteins was tested by comparing the results of test strips with values measured by reference method on analyzer Architect c8000 (Abbott, IL, USA). Precision was estimated by the degree of agreement of repeated measurements, accuracy by the degree of agreement with result of reference method. Comparability was expressed with kappa coefficient with 95% confidence interval. Acceptance criteria were: precision: 90% agreement of repeated measurements, pH: acceptable bias ±1 pH unit, specific gravity: acceptable bias ± 0.005 and comparability: kappa coefficient ≥0.6. MedCalc (v12.7.2.0, Ostend, Belgium) was used for statistical analysis. Results: Precision was acceptable for all parameters, except for leukocytes in one patient›s sample (agreement for only 8/10 measurements). Comparability between analyzers was satisfactory, except for bilirubin (kappa=0.370 (0.112 to 0.628)). Accuracy for bilirubin was satisfactory (agreement 10/10). The observed differences in proteins (8/10) and glucose (6/10) were not clinically significant. Conclusion: Analytical performance of Combur10Test® M on Cobas u411 analyzer are within acceptance limits. Observed bias for glucose and proteins are not clinically significant and are caused by differences in methodology. Analyzers are comparable, except for bilirubin measurement.

Izvorni jezik
Engleski



POVEZANOST RADA


Ustanove:
KBC "Sestre Milosrdnice"

Profili:

Avatar Url Alen Vrtarić (autor)

Avatar Url Ana-Maria Šimundić (autor)

Avatar Url Nora Nikolac (autor)

Avatar Url Marijana Miler (autor)


Citiraj ovu publikaciju:

Vrtarić, Alen; Nikolac, Nora; Miler, Marijana; Šimundić, Ana-Maria
Analytical verification of Combur10 Test® M strips and urine analyzer Cobas u411 // Biochemia Medica / Šimundić, Ana-Maria (ur.).
Zagreb: Hrvatsko društvo za medicinsku biokemiju i laboratorijsku medicinu (HDMBLM), 2015. str. S100-S101 (poster, domaća recenzija, sažetak, znanstveni)
Vrtarić, A., Nikolac, N., Miler, M. & Šimundić, A. (2015) Analytical verification of Combur10 Test® M strips and urine analyzer Cobas u411. U: Šimundić, A. (ur.)Biochemia Medica.
@article{article, author = {Vrtari\'{c}, Alen and Nikolac, Nora and Miler, Marijana and \v{S}imundi\'{c}, Ana-Maria}, editor = {\v{S}imundi\'{c}, A.}, year = {2015}, pages = {S100-S101}, keywords = {method verification, method comparability, precision}, title = {Analytical verification of Combur10 Test® M strips and urine analyzer Cobas u411}, keyword = {method verification, method comparability, precision}, publisher = {Hrvatsko dru\v{s}tvo za medicinsku biokemiju i laboratorijsku medicinu (HDMBLM)}, publisherplace = {Rijeka, Hrvatska} }
@article{article, author = {Vrtari\'{c}, Alen and Nikolac, Nora and Miler, Marijana and \v{S}imundi\'{c}, Ana-Maria}, editor = {\v{S}imundi\'{c}, A.}, year = {2015}, pages = {S100-S101}, keywords = {method verification, method comparability, precision}, title = {Analytical verification of Combur10 Test® M strips and urine analyzer Cobas u411}, keyword = {method verification, method comparability, precision}, publisher = {Hrvatsko dru\v{s}tvo za medicinsku biokemiju i laboratorijsku medicinu (HDMBLM)}, publisherplace = {Rijeka, Hrvatska} }




Contrast
Increase Font
Decrease Font
Dyslexic Font