Pregled bibliografske jedinice broj: 819274

Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes


Juričić, Gordana; Bakliža, Ana; Saračević, Andrea; Milevoj Kopčinović, Lara; Dobrijević, Sanja; Drmić, Sandra; Šimundić, Ana-Maria
Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes // Clinical chemistry and laboratory medicine, 54 (2016), 3; 411-418 doi:10.1515/cclm-2015-0746 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 819274 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes

Autori
Juričić, Gordana ; Bakliža, Ana ; Saračević, Andrea ; Milevoj Kopčinović, Lara ; Dobrijević, Sanja ; Drmić, Sandra ; Šimundić, Ana-Maria

Izvornik
Clinical chemistry and laboratory medicine (1434-6621) 54 (2016), 3; 411-418

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
diabetes ; Glucomedics ; glycolysis ; glycolysis inhibitors ; preanalytical phase

Sažetak
Background: Delayed sample processing can affect accurate glucose measurement. Our aim was to investigate the stability of glucose in samples collected in serum, sodium fluoride/potassium oxalate (NaF/KOx) and Glucomedics tubes processed according to different controlled pre-centrifugation delays (up to 180 min after venipuncture) in order to simulate prolonged sample transport between venipuncture and centrifugation. Methods: Samples were collected from healthy volunteers (n=80) into either serum or NaF/KOx and Glucomedics tubes. Glucose concentration was measured in samples centrifuged immediately after venipuncture and compared with tubes processed with a delay of 60, 120 and 180 min prior to centrifugation. Differences between baseline and respective delayed centrifugation glucose value for each tube type were tested using the paired t-test. Mean bias calculated for each tube type and delay protocol was compared to recommended quality specifications for glucose (2.2%). Results: Glucose concentrations measured in all three delayed tube types were lower in comparison to respective baseline glucose concentrations measured in immediately processed tube (p<0.001). The highest decrease in glucose was observed in serum tubes in all specified time points (p<0.001), while glucose was most stable in Glucomedics tubes (p<0.001). The decrease in glucose observed for serum and NaF/KOx tubes was clinically significant at all specified time points while the bias for Glucomedics tubes did not exceed the criteria even with a centrifugation delay of 180 min. Conclusions: Glucose stability in un- centrifuged Glucomedics tubes is much superior to serum and NaF/KOx tubes. Glucomedics tubes can be left un-centrifuged for up to 3 h without affecting glucose concentration.

Izvorni jezik
Engleski

Znanstvena područja
Kliničke medicinske znanosti



POVEZANOST RADA


Projekti:
134-1340227-0200 - Upala i udio farmakogenetike u razvoju i ishodu akutnih i kroničnih bolesti (Šimundić, Ana-Maria, MZOS ) ( CroRIS)

Ustanove:
KBC "Sestre Milosrdnice"

Poveznice na cjeloviti tekst rada:

doi www.degruyter.com

Citiraj ovu publikaciju:

Juričić, Gordana; Bakliža, Ana; Saračević, Andrea; Milevoj Kopčinović, Lara; Dobrijević, Sanja; Drmić, Sandra; Šimundić, Ana-Maria
Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes // Clinical chemistry and laboratory medicine, 54 (2016), 3; 411-418 doi:10.1515/cclm-2015-0746 (međunarodna recenzija, članak, znanstveni)
Juričić, G., Bakliža, A., Saračević, A., Milevoj Kopčinović, L., Dobrijević, S., Drmić, S. & Šimundić, A. (2016) Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes. Clinical chemistry and laboratory medicine, 54 (3), 411-418 doi:10.1515/cclm-2015-0746.
@article{article, author = {Juri\v{c}i\'{c}, Gordana and Bakli\v{z}a, Ana and Sara\v{c}evi\'{c}, Andrea and Milevoj Kop\v{c}inovi\'{c}, Lara and Dobrijevi\'{c}, Sanja and Drmi\'{c}, Sandra and \v{S}imundi\'{c}, Ana-Maria}, year = {2016}, pages = {411-418}, DOI = {10.1515/cclm-2015-0746}, keywords = {diabetes, Glucomedics, glycolysis, glycolysis inhibitors, preanalytical phase}, journal = {Clinical chemistry and laboratory medicine}, doi = {10.1515/cclm-2015-0746}, volume = {54}, number = {3}, issn = {1434-6621}, title = {Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes}, keyword = {diabetes, Glucomedics, glycolysis, glycolysis inhibitors, preanalytical phase} }
@article{article, author = {Juri\v{c}i\'{c}, Gordana and Bakli\v{z}a, Ana and Sara\v{c}evi\'{c}, Andrea and Milevoj Kop\v{c}inovi\'{c}, Lara and Dobrijevi\'{c}, Sanja and Drmi\'{c}, Sandra and \v{S}imundi\'{c}, Ana-Maria}, year = {2016}, pages = {411-418}, DOI = {10.1515/cclm-2015-0746}, keywords = {diabetes, Glucomedics, glycolysis, glycolysis inhibitors, preanalytical phase}, journal = {Clinical chemistry and laboratory medicine}, doi = {10.1515/cclm-2015-0746}, volume = {54}, number = {3}, issn = {1434-6621}, title = {Glucose is stable during prolonged storage in un-centrifuged Greiner tubes with liquid citrate buffer, but not in serum and NaF/KOx tubes}, keyword = {diabetes, Glucomedics, glycolysis, glycolysis inhibitors, preanalytical phase} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font