Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 793918

Radiation doses on CT scanner in UH Rijeka and comparison with European Commision dose reference levels


Šegota, Doris; Jurković, Slaven; Posedel, Dario; Kralik, Ivana; Miletić, Damir; Faj, Dario
Radiation doses on CT scanner in UH Rijeka and comparison with European Commision dose reference levels // Proceedings of the ninth symposium of the CRPA
Krk, Hrvatska: Hrvatsko društvo za zaštitu od zračenja, 2013. str. 259-264 (predavanje, domaća recenzija, cjeloviti rad (in extenso), stručni)


CROSBI ID: 793918 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Radiation doses on CT scanner in UH Rijeka and comparison with European Commision dose reference levels

Autori
Šegota, Doris ; Jurković, Slaven ; Posedel, Dario ; Kralik, Ivana ; Miletić, Damir ; Faj, Dario

Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), stručni

Izvornik
Proceedings of the ninth symposium of the CRPA / - : Hrvatsko društvo za zaštitu od zračenja, 2013, 259-264

ISBN
978-953-96133-8-7

Skup
Deveti simpozij Hrvatskog društva za zaštitu od zračenja

Mjesto i datum
Krk, Hrvatska, 10.04.2013. - 12.04.2013

Vrsta sudjelovanja
Predavanje

Vrsta recenzije
Domaća recenzija

Ključne riječi
computed tomography ; dosimetry

Sažetak
The purpose of this study was to compare values of Local Diagnostic Reference Levels (LDRLs) of CT examination protocols used in UH Rijeka with European Commission (EC) Diagnostic Reference Dose Levels (DRLs) and to estimate patient effective dose per type of examination. The dosimetric quantities proposed for DRLs in the European Guidelines (EG) for CT are weighted computed tomography dose index (CTDIw) for a single slice and DLP for a complete examination. Patient’s data (weight, height) as well as technical parameters for each type of examination were collected on Siemens Sensation 16 CT scanner installed at Radiology Department, UH Rijeka. Examinations were categorized according to anatomy: brain, chest, abdomen and pelvis. Computed tomography dose index (CTDI) measurements were performed in CTDI phantom to assure readings accuracy. Both values, CTDIw and DLP, for all four examinations were within proposed values given by European Commission. Also, estimations of effective dose to the average patient for every type of CT examination were made. Results confirmed that UH Rijeka meets EC DRLs for all examinations.

Izvorni jezik
Engleski

Znanstvena područja
Temeljne medicinske znanosti, Kliničke medicinske znanosti



POVEZANOST RADA


Profili:

Avatar Url Damir Miletić (autor)

Avatar Url Ivana Kralik (autor)

Avatar Url Doris Šegota (autor)

Avatar Url Slaven Jurković (autor)

Avatar Url Dario Faj (autor)


Citiraj ovu publikaciju:

Šegota, Doris; Jurković, Slaven; Posedel, Dario; Kralik, Ivana; Miletić, Damir; Faj, Dario
Radiation doses on CT scanner in UH Rijeka and comparison with European Commision dose reference levels // Proceedings of the ninth symposium of the CRPA
Krk, Hrvatska: Hrvatsko društvo za zaštitu od zračenja, 2013. str. 259-264 (predavanje, domaća recenzija, cjeloviti rad (in extenso), stručni)
Šegota, D., Jurković, S., Posedel, D., Kralik, I., Miletić, D. & Faj, D. (2013) Radiation doses on CT scanner in UH Rijeka and comparison with European Commision dose reference levels. U: Proceedings of the ninth symposium of the CRPA.
@article{article, author = {\v{S}egota, Doris and Jurkovi\'{c}, Slaven and Posedel, Dario and Kralik, Ivana and Mileti\'{c}, Damir and Faj, Dario}, year = {2013}, pages = {259-264}, keywords = {computed tomography, dosimetry}, isbn = {978-953-96133-8-7}, title = {Radiation doses on CT scanner in UH Rijeka and comparison with European Commision dose reference levels}, keyword = {computed tomography, dosimetry}, publisher = {Hrvatsko dru\v{s}tvo za za\v{s}titu od zra\v{c}enja}, publisherplace = {Krk, Hrvatska} }
@article{article, author = {\v{S}egota, Doris and Jurkovi\'{c}, Slaven and Posedel, Dario and Kralik, Ivana and Mileti\'{c}, Damir and Faj, Dario}, year = {2013}, pages = {259-264}, keywords = {computed tomography, dosimetry}, isbn = {978-953-96133-8-7}, title = {Radiation doses on CT scanner in UH Rijeka and comparison with European Commision dose reference levels}, keyword = {computed tomography, dosimetry}, publisher = {Hrvatsko dru\v{s}tvo za za\v{s}titu od zra\v{c}enja}, publisherplace = {Krk, Hrvatska} }




Contrast
Increase Font
Decrease Font
Dyslexic Font