Verification of the treatment planning systems in the Croatian radiotherapy community


Švabić, Manda; Jurković, Slaven; Diklić, Ana; Smilović Radojčić, Đeni; Dundara, Dea; Kasabašić, Mladen; Ivković, Ana; Faj, Dario
Verification of the treatment planning systems in the Croatian radiotherapy community // Abstract book, 5th Alpe-Adria Medical Physics Meeting, Trieste, 2012.
Trst, Italija, 2012. (predavanje, međunarodna recenzija, sažetak, znanstveni)


CROSBI ID: 793121 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Verification of the treatment planning systems in the Croatian radiotherapy community

Autori
Švabić, Manda ; Jurković, Slaven ; Diklić, Ana ; Smilović Radojčić, Đeni ; Dundara, Dea ; Kasabašić, Mladen ; Ivković, Ana ; Faj, Dario

Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni

Izvornik
Abstract book, 5th Alpe-Adria Medical Physics Meeting, Trieste, 2012. / - , 2012

Skup
5th Alpe-Adria Medical Physics Meeting

Mjesto i datum
Trst, Italija, 03.05.2012. - 06.05.2012

Vrsta sudjelovanja
Predavanje

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
radiotherapy ; treatment planning system ; quality assurance

Sažetak
Effective programme for quality assurance (QA) and quality control (QC) depends strongly on the understanding the limitations of various systems used in radiation therapy. Besides the mechanical and dosimetrical parameters of therapy machines [1], the treatment planning system (TPS) accuracy has a great impact on the treatment outcome. The verification of TPS was done using homogenous (water) phantom and anthropomorphic phantom CIRS Thorax (CIRS Inc., Norfolk). The first step was to check calibration and beam modelling in water through 58 different tests. These tests were planned in CMS XiO for two photon energies and multigrid superposition algorithm. The criteria were set according to IAEA TRS 430 [2]. After the verification in water, measurements were performed in anthropomorphic phantom. New set of 14 tests were also planned in XiO for two photon energies and multigrid superposition algorithm. The same dosimetrical verification of TPS was done in 4 radiotherapy centers in Croatia. The results of the measurements in water equivalent tissue were mostly in tolerance. Measurements done in non-water equivalent tissue revealed a significant amount of out of tolerance values, especially for high energy photon beams. Discrepancies between measurement and calculation may arise due to a number of factors [3]. After checking most of them, it becomes apparent that the limitation of calculating system cannot be neglected and the doses calculated in inhomogeneous volumes should be additionally reconsidered. Bibliography 1. Croatian Medical and Biological Engineering Society / Medical Physics Division. Prirunik za kontrolu kvalitete linearnog akceleratora [Linear accelerator quality control handbook, in Croatian] [displayed 25 July 2011]. Available at http://medfizika.netai.net/upload/Protokol_Onco rOsijek Rijeka.doc 2. IAEA (International Atomic Energy Agency) Technical Report Series 430 Commissioning and quality assurance of computerized planning systems for radiation treatment of cancer. Vienna: IAEA 2005. 3. Gershkevitsh E, Schmidt R, Velez G et al. Dosimetric verification of radiotherapy treatment planning systems: Results of IAEA pilot study. Radioth Oncol 2009 ; 89:338-346

Izvorni jezik
Engleski

Znanstvena područja
Fizika



POVEZANOST RADA



Citiraj ovu publikaciju:

Švabić, Manda; Jurković, Slaven; Diklić, Ana; Smilović Radojčić, Đeni; Dundara, Dea; Kasabašić, Mladen; Ivković, Ana; Faj, Dario
Verification of the treatment planning systems in the Croatian radiotherapy community // Abstract book, 5th Alpe-Adria Medical Physics Meeting, Trieste, 2012.
Trst, Italija, 2012. (predavanje, međunarodna recenzija, sažetak, znanstveni)
Švabić, M., Jurković, S., Diklić, A., Smilović Radojčić, Đ., Dundara, D., Kasabašić, M., Ivković, A. & Faj, D. (2012) Verification of the treatment planning systems in the Croatian radiotherapy community. U: Abstract book, 5th Alpe-Adria Medical Physics Meeting, Trieste, 2012..
@article{article, author = {\v{S}vabi\'{c}, Manda and Jurkovi\'{c}, Slaven and Dikli\'{c}, Ana and Smilovi\'{c} Radoj\v{c}i\'{c}, \DJeni and Dundara, Dea and Kasaba\v{s}i\'{c}, Mladen and Ivkovi\'{c}, Ana and Faj, Dario}, year = {2012}, keywords = {radiotherapy, treatment planning system, quality assurance}, title = {Verification of the treatment planning systems in the Croatian radiotherapy community}, keyword = {radiotherapy, treatment planning system, quality assurance}, publisherplace = {Trst, Italija} }
@article{article, author = {\v{S}vabi\'{c}, Manda and Jurkovi\'{c}, Slaven and Dikli\'{c}, Ana and Smilovi\'{c} Radoj\v{c}i\'{c}, \DJeni and Dundara, Dea and Kasaba\v{s}i\'{c}, Mladen and Ivkovi\'{c}, Ana and Faj, Dario}, year = {2012}, keywords = {radiotherapy, treatment planning system, quality assurance}, title = {Verification of the treatment planning systems in the Croatian radiotherapy community}, keyword = {radiotherapy, treatment planning system, quality assurance}, publisherplace = {Trst, Italija} }




Contrast
Increase Font
Decrease Font
Dyslexic Font