Optimization of 90Sr screening method in waters via Cherenkov radiation


Stojković, Ivana; Todorović, Nataša; Nikolov, Jovana; Krajcar Bronić, Ines; Borković, Damir; Barešić, Jadranka; Sironić, Andreja
Optimization of 90Sr screening method in waters via Cherenkov radiation // Zbornik radova Desetog simpozija Hrvatskog društva za zaštitu od zračenja / Petrinec, Branko ; Bituh, Tomislav ; Milić, Mirta ; Kopjar, Nevenka (ur.).
Zagreb: Hrvatsko društvo za zaštitu od zračenja, 2015. str. 372-377 (poster, nije recenziran, cjeloviti rad (in extenso), znanstveni)


CROSBI ID: 758790 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Optimization of 90Sr screening method in waters via Cherenkov radiation

Autori
Stojković, Ivana ; Todorović, Nataša ; Nikolov, Jovana ; Krajcar Bronić, Ines ; Borković, Damir ; Barešić, Jadranka ; Sironić, Andreja

Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni

Izvornik
Zbornik radova Desetog simpozija Hrvatskog društva za zaštitu od zračenja / Petrinec, Branko ; Bituh, Tomislav ; Milić, Mirta ; Kopjar, Nevenka - Zagreb : Hrvatsko društvo za zaštitu od zračenja, 2015, 372-377

Skup
Deseti simpozij Hrvatskog društva za zaštitu od zračenja

Mjesto i datum
Šibenik, Hrvatska, 15.04.2015. - 17.04.2015

Vrsta sudjelovanja
Poster

Vrsta recenzije
Nije recenziran

Ključne riječi
90Sr; water; screening; LSC; Cherenkov

Sažetak
Rapid screening of strontium activity in waters via Cherenkov radiation detection by ultra-low-level liquid scintillation counter Quantulus 1220 was established in two laboratories, Laboratory for low radioactivity of the Department of Physics, University of Novi Sad, Serbia (UNS) and Laboratory for measurements of low-level radioactivities at the Ruđer Bošković Institute, Zagreb, Croatia (RBI). Optimization of the method comprises choice of an appropriate vial type, selection of optimal spectral window, determination of the counting efficiency in various vials, evaluation of the influence of vial on efficiency and background. The efficiency of the measurement of A(90Sr) in plastic vials is 45.6 % and 37.3 % at UNS and RBI, respectively, while in glass vials it is lower (~32 %). Background count rate is lower in plastic than in glass vials, resulting in better counting properties of the plastic vials. The achieved minimum detectable activity for a total counting time of 300 min was 0.32 Bq∙L-1 in plastic vials at UNS and 0.34 Bq L-1 at RBI. Validation of the method was performed by determination of strontium level in spiked samples within interlaboratory comparison measurements. The A(90Sr) values determined in both laboratories did not deviate significantly from the spiked activities, giving the z-scores in the interval [-2, 2]. Only one sample at RBI had the z-score higher than 2 but lower than 3. All results from UNS had positive z-score values, including the sample measured within the IAEA proficiency test. The 90Sr/90Y Cherenkov radiation measurements of water samples can be successfully used for rapid screening of samples without any pretreatment of the samples.

Izvorni jezik
Engleski

Znanstvena područja
Fizika, Kemija



POVEZANOST RADA


Ustanove:
Institut "Ruđer Bošković", Zagreb

Poveznice na cjeloviti tekst rada:

Pristup cjelovitom tekstu rada www.hdzz.hr

Citiraj ovu publikaciju:

Stojković, Ivana; Todorović, Nataša; Nikolov, Jovana; Krajcar Bronić, Ines; Borković, Damir; Barešić, Jadranka; Sironić, Andreja
Optimization of 90Sr screening method in waters via Cherenkov radiation // Zbornik radova Desetog simpozija Hrvatskog društva za zaštitu od zračenja / Petrinec, Branko ; Bituh, Tomislav ; Milić, Mirta ; Kopjar, Nevenka (ur.).
Zagreb: Hrvatsko društvo za zaštitu od zračenja, 2015. str. 372-377 (poster, nije recenziran, cjeloviti rad (in extenso), znanstveni)
Stojković, I., Todorović, N., Nikolov, J., Krajcar Bronić, I., Borković, D., Barešić, J. & Sironić, A. (2015) Optimization of 90Sr screening method in waters via Cherenkov radiation. U: Petrinec, B., Bituh, T., Milić, M. & Kopjar, N. (ur.)Zbornik radova Desetog simpozija Hrvatskog društva za zaštitu od zračenja.
@article{article, author = {Stojkovi\'{c}, Ivana and Todorovi\'{c}, Nata\v{s}a and Nikolov, Jovana and Krajcar Broni\'{c}, Ines and Borkovi\'{c}, Damir and Bare\v{s}i\'{c}, Jadranka and Sironi\'{c}, Andreja}, year = {2015}, pages = {372-377}, keywords = {90Sr, water, screening, LSC, Cherenkov}, title = {Optimization of 90Sr screening method in waters via Cherenkov radiation}, keyword = {90Sr, water, screening, LSC, Cherenkov}, publisher = {Hrvatsko dru\v{s}tvo za za\v{s}titu od zra\v{c}enja}, publisherplace = {\v{S}ibenik, Hrvatska} }
@article{article, author = {Stojkovi\'{c}, Ivana and Todorovi\'{c}, Nata\v{s}a and Nikolov, Jovana and Krajcar Broni\'{c}, Ines and Borkovi\'{c}, Damir and Bare\v{s}i\'{c}, Jadranka and Sironi\'{c}, Andreja}, year = {2015}, pages = {372-377}, keywords = {90Sr, water, screening, LSC, Cherenkov}, title = {Optimization of 90Sr screening method in waters via Cherenkov radiation}, keyword = {90Sr, water, screening, LSC, Cherenkov}, publisher = {Hrvatsko dru\v{s}tvo za za\v{s}titu od zra\v{c}enja}, publisherplace = {\v{S}ibenik, Hrvatska} }




Contrast
Increase Font
Decrease Font
Dyslexic Font