Pregled bibliografske jedinice broj: 698900
Noise in ultrasonic imaging
Noise in ultrasonic imaging // Ultrasonics, 18 (1980), 2; 81-84 doi:10.1016/0041-624X(80)90141-9 (podatak o recenziji nije dostupan, članak, ostalo)
CROSBI ID: 698900 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Noise in ultrasonic imaging
Autori
Breyer, Branko ; Viculin, Tomislav ; Vojnović, Božidar
Izvornik
Ultrasonics (0041-624X) 18
(1980), 2;
81-84
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, ostalo
Ključne riječi
Noise ; ultrasonic imaging
Sažetak
The feasibility of noise measurement in medical ultrasonic imaging has been studied and a theoretical evaluation of stochastic noise influence upon the perceived image has been undertaken. The influence of noise on a perceived image has been calculated by considering the light intensity perception characteristic of the human eye and it has been found that the additive noise in the video signal pushes the small signals in the image towards the invisible (black) region. Impulse noise measurements of amplitude and time jitter have been set up using a standard water/C Cl(4) interface and pulse amplitude analysis to take into account the random nature of noise by distribution measurements.
Izvorni jezik
Engleski
Znanstvena područja
Fizika, Kliničke medicinske znanosti
POVEZANOST RADA
Ustanove:
Klinika za tumore,
Institut "Ruđer Bošković", Zagreb,
Klinički bolnički centar Zagreb
Citiraj ovu publikaciju:
Časopis indeksira:
- Scopus
- MEDLINE