Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 687202

Optimization of low-level LS counter Quantulus 1220 for tritium determination in water samples


Jakonić, Ivana; Todorović, Nataša; Nikolov, Jovana; Krajcar Bronić, Ines; Tenjović, Branislava; Vesković, Miroslav
Optimization of low-level LS counter Quantulus 1220 for tritium determination in water samples // Radiation physics and chemistry (1993), 98 (2014), 69-76 doi:10.1016/j.radphyschem.2014.01.012 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 687202 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Optimization of low-level LS counter Quantulus 1220 for tritium determination in water samples

Autori
Jakonić, Ivana ; Todorović, Nataša ; Nikolov, Jovana ; Krajcar Bronić, Ines ; Tenjović, Branislava ; Vesković, Miroslav

Izvornik
Radiation physics and chemistry (1993) (0969-806X) 98 (2014); 69-76

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
tritium measurement; liquid scintillation counting; scintillation cocktails; optimization

Sažetak
Liquid scintillation counting (LSC) is the most commonly used technique for measuring tritium. To optimize tritium analysis in waters by ultra-low background liquid scintillation spectrometer Quantulus 1220 the optimization of sample/scintillant ratio, choice of appropriate scintillation cocktail and comparison of their efficiency, background and minimal detectable activity (MDA), the effect of chemi- and photoluminescence and combination of scintillant/vial were performed. ASTM D4107-08 [1] method had been successfully applied in our laboratory for two years. During our last preparation of samples a serious quench effect in count rates of samples that could be consequence of possible contamination by DMSO was noticed. The goal of this paper is to demonstrate development of new direct method in our laboratory proposed by Pujol and Sanchez-Cabeza [2], which turned out to be faster and simpler than ASTM method while we are dealing with problem of neutralization of DMSO in apparatus. The minimum detectable activity achieved was 2.0 Bq•1-1 for a total counting time of 300 minutes. In order to test the optimization of system for this method tritium level was determined in Danube river samples and also for several samples within intercomparison with Ruđer Bošković Institute (IRB).

Izvorni jezik
Engleski

Znanstvena područja
Fizika, Geologija



POVEZANOST RADA


Projekti:
098-0982709-2741 - Prirodni radioizotopi u istraživanju krških ekosustava i datiranju (Horvatinčić, Nada, MZOS ) ( CroRIS)

Ustanove:
Institut "Ruđer Bošković", Zagreb

Profili:

Avatar Url Ines Krajcar Bronić (autor)

Poveznice na cjeloviti tekst rada:

doi www.sciencedirect.com

Citiraj ovu publikaciju:

Jakonić, Ivana; Todorović, Nataša; Nikolov, Jovana; Krajcar Bronić, Ines; Tenjović, Branislava; Vesković, Miroslav
Optimization of low-level LS counter Quantulus 1220 for tritium determination in water samples // Radiation physics and chemistry (1993), 98 (2014), 69-76 doi:10.1016/j.radphyschem.2014.01.012 (međunarodna recenzija, članak, znanstveni)
Jakonić, I., Todorović, N., Nikolov, J., Krajcar Bronić, I., Tenjović, B. & Vesković, M. (2014) Optimization of low-level LS counter Quantulus 1220 for tritium determination in water samples. Radiation physics and chemistry (1993), 98, 69-76 doi:10.1016/j.radphyschem.2014.01.012.
@article{article, author = {Jakoni\'{c}, Ivana and Todorovi\'{c}, Nata\v{s}a and Nikolov, Jovana and Krajcar Broni\'{c}, Ines and Tenjovi\'{c}, Branislava and Veskovi\'{c}, Miroslav}, year = {2014}, pages = {69-76}, DOI = {10.1016/j.radphyschem.2014.01.012}, keywords = {tritium measurement, liquid scintillation counting, scintillation cocktails, optimization}, journal = {Radiation physics and chemistry (1993)}, doi = {10.1016/j.radphyschem.2014.01.012}, volume = {98}, issn = {0969-806X}, title = {Optimization of low-level LS counter Quantulus 1220 for tritium determination in water samples}, keyword = {tritium measurement, liquid scintillation counting, scintillation cocktails, optimization} }
@article{article, author = {Jakoni\'{c}, Ivana and Todorovi\'{c}, Nata\v{s}a and Nikolov, Jovana and Krajcar Broni\'{c}, Ines and Tenjovi\'{c}, Branislava and Veskovi\'{c}, Miroslav}, year = {2014}, pages = {69-76}, DOI = {10.1016/j.radphyschem.2014.01.012}, keywords = {tritium measurement, liquid scintillation counting, scintillation cocktails, optimization}, journal = {Radiation physics and chemistry (1993)}, doi = {10.1016/j.radphyschem.2014.01.012}, volume = {98}, issn = {0969-806X}, title = {Optimization of low-level LS counter Quantulus 1220 for tritium determination in water samples}, keyword = {tritium measurement, liquid scintillation counting, scintillation cocktails, optimization} }