Intracardiac electrogram method for atrioventricular and interventricular delay optimization: comparibility to standard echocardiographic method
Intracardiac electrogram method for atrioventricular and interventricular delay optimization: comparibility to standard echocardiographic method // EUROECHO 2011 European journal 2chocardiogr 12 (2011) (S2) ii90-ii120
Budimpešta, Mađarska, 2011. str. ii90-ii120 (poster, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 669323 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Intracardiac electrogram method for atrioventricular and interventricular delay optimization: comparibility to standard echocardiographic method
Autori
Pezo Nikolić, Borka ; Jurin, Hrvoje ; Lovrić, Danijel ; Baričević, Željko ; Ivanac Vranešić, Irena ; Lovrić Benčič, Martina ; Ernst, Aleksander ; Šeparović Hanževački, Jadranka
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
EUROECHO 2011 European journal 2chocardiogr 12 (2011) (S2) ii90-ii120
/ - , 2011, Ii90-ii120
Skup
EUROECHO 2011
Mjesto i datum
Budimpešta, Mađarska, 07.12.2011. - 10.12.2011
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
cardiac resyncronization ; optimization of syncronization therapy
Sažetak
Abstract Background: Studies have shown that echocardiographic (ECHO) optimization of atrioventricular (AV) and interventricular (VV) delays in cardiac resyncronization therapy (CRT) patients results in improvement of cardiac output by reducing left ventricular (LV) dyssynchrony. Intracardiac electrogram method (IEGM) is a novel technique based on the analysis of atrial intrinsic depolarization and interventricular conduction delay with yet undetermined clinical value. The aim of this study was to determine the efficacy of the IEGM compared to conventional ECHO-guided CRT optimization. Methods: 8 patients previously implanted with Medtronic CRT (3F/5M, 62.5±10.5 yrs) underwent IEGM-based delay optimization with optimal AV and VV delays obtained. Subsequently, MR dp/dt and LVOT VTI as echocardiographic markers of resynchronization effectiveness were measured. ECHO-guided optimization of the AV and VV dyssynchrony followed using PW Doppler of the mitral inflow and LV/RV outflow tract to measure interventricular mechanical delay, respectively. Pearson correlation analysis was used to evaluate findings. Results: A strong correlation between ECHO and IEGM-guided measurements was found, as follows: LVOT VTI (cc=0.979, p=0.896), dp/dt (cc =0.993 p=0.977), the final AV (cc=0.948, p=0.967) and VV delay (cc=0.980, p=0, 787). However, the statistical significance was not achieved due to small studied sample. Conclusion: The IEGM proved to be as effective as the standard ECHO CRT optimization with the advantage of being less- time consuming. Due to present lack of scientific evidence further studies are required.⇓ Figure View larger version: In this page In a new window Download as PowerPoint Slide Published on behalf of the European Society of Cardiology. All rights reserved. © The Author 2011. For permissions please email: journals.permissions@oup.com
Izvorni jezik
Engleski
Napomena
Rad je kao poster prezentiran na skupu EUROECHO 2011 održanom od 07.-10.12.2011.g., Budimpešta, Mađarska. European journal 2chocardiogr 12 (2011) (S2) ii90-ii120
POVEZANOST RADA
Projekti:
108-1081875-1991 - Doppler miokarda u ranom otkrivanju i praćenju kardiovaskularnih bolesti (Šeparović-Hanževački, Jadranka, MZOS ) ( CroRIS)
Ustanove:
Medicinski fakultet, Zagreb
Profili:
Borka Pezo Nikolić
(autor)
Jadranka Šeparović-Hanževački
(autor)
Irena Ivanac Vranešić
(autor)
Martina Lovrić Benčić
(autor)
Aleksander Ernst
(autor)