Comparative analysis of the performances of Abbott Architect c8000 and Roche Hitachi 917 for 25 analytes in daily routine laboratory workflow


Dujmov, Ivana; Čepić, Katarina; Šupe-Domić, Daniela; Mladina, Blanka; Ivanović-Ivelja, Nada
Comparative analysis of the performances of Abbott Architect c8000 and Roche Hitachi 917 for 25 analytes in daily routine laboratory workflow // Clinical chemistry and laboratory medicine / Gerard Siest, Nancy (ur.).
Berlin: International Federation of Clinical Chemistry and Laboratory Medicine, 2004. str. 70-71 (poster, međunarodna recenzija, sažetak, stručni)


CROSBI ID: 665123 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Comparative analysis of the performances of Abbott Architect c8000 and Roche Hitachi 917 for 25 analytes in daily routine laboratory workflow

Autori
Dujmov, Ivana ; Čepić, Katarina ; Šupe-Domić, Daniela ; Mladina, Blanka ; Ivanović-Ivelja, Nada

Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, stručni

Izvornik
Clinical chemistry and laboratory medicine / Gerard Siest, Nancy - Berlin : International Federation of Clinical Chemistry and Laboratory Medicine, 2004, 70-71

Skup
8th Alps Adria Congress of Clinical Chemistry and Laboratory Medicine

Mjesto i datum
Luzern, Švicarska, 05.10.2004. - 09.10.2004

Vrsta sudjelovanja
Poster

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
Evaluation study ; Abbott Architect c8000 ; Roche Hitachi 917 ; different analytes

Sažetak
Modern analytical methods in laboratory include highly reliable clinical chemistry systems.For each new clinical analyzer in laboratory we performed the evaluation study. In this study we compared the Abbott Architect c8000 and Roche Hitachi 917 instruments for 25 different analytes:10 substrates, 8 enzymes and 7 electrolytes. The Abbott Architect c8000 and Roche Hitachi 917 systems are fully automated, random Access, clinical chemistry workcells for the rapid quantitative analysis of tests commonly performed in clinical chemistry section of the laboratory. Analytes examined included 25 different analytes:glucose, urea, creatinine, uric acid, bilirubin total, bilirubin direct, total protein, albumine, cholesterol, triglyceride, Na, K, Cl, calcium, inorganic phosphorus, magnesium, iron, LDH, AST, ALT, GGT, CK, CKMB, alcaline phosphatase, alfa-amylase. Coefficients of correlation values obtained by linear regression analysis were ranged from 0.8745(slightly different values of only sodium ion concentrations)to 0.9995. Regression method showed comparable results between the Abbott Architect c8000 system and the Roche Hitachi 917 analyzer.Based on these results, the Abbott Architect c8000 analyzer demonstrates excellent analytical performance characteristics and good concordance with the Roche Hitachi 917 analyzer.

Izvorni jezik
Engleski

Znanstvena područja
Temeljne medicinske znanosti



POVEZANOST RADA


Ustanove:
KBC Split

Profili:

Avatar Url Daniela Šupe Domić (autor)


Citiraj ovu publikaciju:

Dujmov, Ivana; Čepić, Katarina; Šupe-Domić, Daniela; Mladina, Blanka; Ivanović-Ivelja, Nada
Comparative analysis of the performances of Abbott Architect c8000 and Roche Hitachi 917 for 25 analytes in daily routine laboratory workflow // Clinical chemistry and laboratory medicine / Gerard Siest, Nancy (ur.).
Berlin: International Federation of Clinical Chemistry and Laboratory Medicine, 2004. str. 70-71 (poster, međunarodna recenzija, sažetak, stručni)
Dujmov, I., Čepić, K., Šupe-Domić, D., Mladina, B. & Ivanović-Ivelja, N. (2004) Comparative analysis of the performances of Abbott Architect c8000 and Roche Hitachi 917 for 25 analytes in daily routine laboratory workflow. U: Gerard Siest, N. (ur.)Clinical chemistry and laboratory medicine.
@article{article, author = {Dujmov, Ivana and \v{C}epi\'{c}, Katarina and \v{S}upe-Domi\'{c}, Daniela and Mladina, Blanka and Ivanovi\'{c}-Ivelja, Nada}, editor = {Gerard Siest, N.}, year = {2004}, pages = {70-71}, keywords = {Evaluation study, Abbott Architect c8000, Roche Hitachi 917, different analytes}, title = {Comparative analysis of the performances of Abbott Architect c8000 and Roche Hitachi 917 for 25 analytes in daily routine laboratory workflow}, keyword = {Evaluation study, Abbott Architect c8000, Roche Hitachi 917, different analytes}, publisher = {International Federation of Clinical Chemistry and Laboratory Medicine}, publisherplace = {Luzern, \v{S}vicarska} }
@article{article, author = {Dujmov, Ivana and \v{C}epi\'{c}, Katarina and \v{S}upe-Domi\'{c}, Daniela and Mladina, Blanka and Ivanovi\'{c}-Ivelja, Nada}, editor = {Gerard Siest, N.}, year = {2004}, pages = {70-71}, keywords = {Evaluation study, Abbott Architect c8000, Roche Hitachi 917, different analytes}, title = {Comparative analysis of the performances of Abbott Architect c8000 and Roche Hitachi 917 for 25 analytes in daily routine laboratory workflow}, keyword = {Evaluation study, Abbott Architect c8000, Roche Hitachi 917, different analytes}, publisher = {International Federation of Clinical Chemistry and Laboratory Medicine}, publisherplace = {Luzern, \v{S}vicarska} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE





Contrast
Increase Font
Decrease Font
Dyslexic Font