Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 664852

Novel external fixation fracture method with circular locking mechanism compared with the application of dynamic axial external fixator on experimental tibial model ensures better stability in bending and favourable performance in dynamic loads


Pavić, Arsen; Kodvanj, Janoš; Sabalić, Srećko; Čukelj, Fabijan; Bakota, Bore
Novel external fixation fracture method with circular locking mechanism compared with the application of dynamic axial external fixator on experimental tibial model ensures better stability in bending and favourable performance in dynamic loads // Injury, 44 (2013), S3; S52-S55 doi:10.1016/S0020-1383(13)70199-2 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 664852 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Novel external fixation fracture method with circular locking mechanism compared with the application of dynamic axial external fixator on experimental tibial model ensures better stability in bending and favourable performance in dynamic loads

Autori
Pavić, Arsen ; Kodvanj, Janoš ; Sabalić, Srećko ; Čukelj, Fabijan ; Bakota, Bore

Izvornik
Injury (0020-1383) 44 (2013), S3; S52-S55

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
tibial fracture; external fixator; novel external fixator; dynamic axial external fixator; dynamic load; static load; cyclic load; biomechanics; Croatia

Sažetak
The aim of this study was to compare the biomechanical properties of a novel tibial external bone fracture fixator with a circular locking mechanism with standard dynamic axial external fixator. In order to investigate the prototype usability in experimental conditions, a biomechanical study was performed in which 42 polyacetal tubes set in 14 experimental groups and subgroups represented the fractured tibia that were fixed by a standard dynamic axial external fixator and a novel fixator. Displacements under static and dynamic loads were measured, with static ones corresponding to three directions of fragment movement and dynamic simulating the human gait. Analysis was performed in SPSS v13, with significance set at P<0.05. The novel fixator showed biomechanical superiority in “fragments apart” study groups, while the standard dynamic axial external fixator outperformed the novel one in the situations of bending with “fragments in contact” study groups. There were no significant differences in dynamic load, despite better numerical result of the novel fixator. The novel fixator is expectedly faster applicable and offers greater extent of external fixation flexibility. Further developments of this model thus seems justified in both construction improvement and on clinical application.

Izvorni jezik
Engleski

Znanstvena područja
Strojarstvo, Temeljne medicinske znanosti



POVEZANOST RADA


Projekti:
120-1201910-1907 - Eksperimentalna simulacija oštećenja mehaničkih i biomehaničkih sustava (Kodvanj, Janoš, MZOS ) ( CroRIS)

Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb,
KBC "Sestre Milosrdnice",
KBC Split,
Klinika za traumatologiju,
Opća bolnica Karlovac

Profili:

Avatar Url Fabijan Čukelj (autor)

Avatar Url Janoš Kodvanj (autor)

Avatar Url Srećko Sabalić (autor)

Poveznice na cjeloviti tekst rada:

doi www.sciencedirect.com dx.doi.org

Citiraj ovu publikaciju:

Pavić, Arsen; Kodvanj, Janoš; Sabalić, Srećko; Čukelj, Fabijan; Bakota, Bore
Novel external fixation fracture method with circular locking mechanism compared with the application of dynamic axial external fixator on experimental tibial model ensures better stability in bending and favourable performance in dynamic loads // Injury, 44 (2013), S3; S52-S55 doi:10.1016/S0020-1383(13)70199-2 (međunarodna recenzija, članak, znanstveni)
Pavić, A., Kodvanj, J., Sabalić, S., Čukelj, F. & Bakota, B. (2013) Novel external fixation fracture method with circular locking mechanism compared with the application of dynamic axial external fixator on experimental tibial model ensures better stability in bending and favourable performance in dynamic loads. Injury, 44 (S3), S52-S55 doi:10.1016/S0020-1383(13)70199-2.
@article{article, author = {Pavi\'{c}, Arsen and Kodvanj, Jano\v{s} and Sabali\'{c}, Sre\'{c}ko and \v{C}ukelj, Fabijan and Bakota, Bore}, year = {2013}, pages = {S52-S55}, DOI = {10.1016/S0020-1383(13)70199-2}, keywords = {tibial fracture, external fixator, novel external fixator, dynamic axial external fixator, dynamic load, static load, cyclic load, biomechanics, Croatia}, journal = {Injury}, doi = {10.1016/S0020-1383(13)70199-2}, volume = {44}, number = {S3}, issn = {0020-1383}, title = {Novel external fixation fracture method with circular locking mechanism compared with the application of dynamic axial external fixator on experimental tibial model ensures better stability in bending and favourable performance in dynamic loads}, keyword = {tibial fracture, external fixator, novel external fixator, dynamic axial external fixator, dynamic load, static load, cyclic load, biomechanics, Croatia} }
@article{article, author = {Pavi\'{c}, Arsen and Kodvanj, Jano\v{s} and Sabali\'{c}, Sre\'{c}ko and \v{C}ukelj, Fabijan and Bakota, Bore}, year = {2013}, pages = {S52-S55}, DOI = {10.1016/S0020-1383(13)70199-2}, keywords = {tibial fracture, external fixator, novel external fixator, dynamic axial external fixator, dynamic load, static load, cyclic load, biomechanics, Croatia}, journal = {Injury}, doi = {10.1016/S0020-1383(13)70199-2}, volume = {44}, number = {S3}, issn = {0020-1383}, title = {Novel external fixation fracture method with circular locking mechanism compared with the application of dynamic axial external fixator on experimental tibial model ensures better stability in bending and favourable performance in dynamic loads}, keyword = {tibial fracture, external fixator, novel external fixator, dynamic axial external fixator, dynamic load, static load, cyclic load, biomechanics, Croatia} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE


Uključenost u ostale bibliografske baze podataka::


  • EMBASE (Excerpta Medica)
  • MEDLINE
  • PubMed
  • Scopus


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font