Pregled bibliografske jedinice broj: 63963
Evaluation of the NOVA-10 STAT/routine chemistry analyzer
Evaluation of the NOVA-10 STAT/routine chemistry analyzer // Clinical Chemistry and Laboratory Medicine ; vol 37:Abstracts volume, Special supplement / Siest, Gerard (ur.).
Milano: Walter de Gruyter, 1999. str. S380-S380 (poster, međunarodna recenzija, sažetak, znanstveni)
CROSBI ID: 63963 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Evaluation of the NOVA-10 STAT/routine chemistry analyzer
Autori
Ferenec-Ružić, Dragica ; Vrkić, Nada ; Blazinić, Franciska ; Lazić, N
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
Clinical Chemistry and Laboratory Medicine ; vol 37:Abstracts volume, Special supplement
/ Siest, Gerard - Milano : Walter de Gruyter, 1999, S380-S380
Skup
IFCC-World Lab, 17th International and 13th European Congress of Clinical Chemistry and Laboratory Medicine
Mjesto i datum
Firenca, Italija, 06.06.1999. - 11.06.1999
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
NOVA-10; electrolyte analyzer; evaluation
Sažetak
The Nova 10 (NOVA Biomedical, Waltham, MA) is an electrolyte analyzer for serum, plasma, CSF and urine testing. It uses a microcomputer-based operating system to measure analytes and calculate both automated and stat-analyses.The aim was to evaluate analytical performance of the instrument according to the ECCLS guidelines (imprecision, accuracy, carry-over, linearity, and comparison studies). Sodium, potassium, chloride, total calcium (tCa) and lithium were evaluated. Control material used in the study Stat Profile Control 1-3, Precinorm U, Precipath U and Serodos. Randomly chosen patient sera were used in comparison study for Na+, K+, Cl- and tCa on the Synchron CX-3 (Beckman), and Li+ on Flame Photometer PHF-104 (Hycel, France). Intra-assay CV was <1% for Na+, K+, Cl-, tCa, and 1.08% for Li+. Inter-assay CV for Na+, K+, Cl-, tCa and Li+ ranged from 1.25%-1.68% respectively. Carry-over was negligible (K<0.01%). Linearity for all analytes was within the declared analytical range. Results of the comparison study (Passing Bablok) showed no statistically significant difference neither for Na+, K+, Cl- and tCa compared with Synchron CX3 (p>0.001), nor for Li+ compared with Flame Photometer. In conclusion, results of the evaluation suggest that NOVA 10 is a precise, accurate and reliable instrument for daily routine.
Izvorni jezik
Engleski
Znanstvena područja
Temeljne medicinske znanosti, Farmacija
POVEZANOST RADA
Projekti:
134003
Ustanove:
Farmaceutsko-biokemijski fakultet, Zagreb,
KBC "Sestre Milosrdnice"
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE