Pregled bibliografske jedinice broj: 636772
Clinical simulations of prostate radiotherapy using BOMAB-like phantoms: results for photons
Clinical simulations of prostate radiotherapy using BOMAB-like phantoms: results for photons // Radiation measurements, 57 (2013), 35-47 doi:10.1016/j.radmeas.2012.12.012 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 636772 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Clinical simulations of prostate radiotherapy using BOMAB-like phantoms: results for photons
Autori
Miljanić, Saveta ; Bessieres, Igor ; Bordy, Jean-Marc ; d’Errico, Francesko ; Di Fulvio, Angela ; Kabat, Damian ; Knežević, Željka ; Olko, Pawel ; Stolarczyk, Liliana ; Tana, Luigi ; Harrison, Roger
Izvornik
Radiation measurements (1350-4487) 57
(2013);
35-47
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
out-of-field doses in radiation therapy; RPL; TLD; OSL dosemeters; IMRT; VMAT; tomotherapy
Sažetak
In this work, water tank experiments were extended to a BOMAB-like phantom. This phantom is more clinically realistic than a water tank, as it is composed of body, legs, arms and head sections in the form of water tanks of circular or elliptical cross section. It is sufficiently realistic to allow the simulation of some clinical treatments. In the experiments to be described, four types of prostate treatment were simulated: (i) Volumetric Modulated arc therapy (VMAT, 6 MV). (ii) Tomotherapy (6 MV) (iii) IMRT (6MV and 18 MV) (iv) 5-field conformal RT (15 MV) and 4-field conformal RT (6 MV and 18 MV Doses were measured in the BOMAB phantom over an axial distance of approximately 50 cm, at positions along five pipes, one on the longitudinal axis of the phantom and the remaining four adjacent and parallel. TLD, OSL and RPL dosemeters were used and a comparision of dosemeters under the same irradiation conditions showed that dosemeters generally agreed to within 3% compared with ion chamber reference measurements. Other comparisons were possible with this data. They include a comparison of doses (beam profiles) in different positions in the BOMAB phantom, a comparison of different treatment modalities in the two contributing clinical centres (Pisa and Krakow) and a comparison of dose profiles resulting from the different treatment techniques and the corresponding doses calculated by the treatment planning systems used to generate the treatment plans. Finally, preliminary measurements of surface doses at selected points on the trunk of the BOMAB phantom were made using TLD and diode detectors. The aim of this work is to investigate the possible relationships between surface doses and underlying doses within the phantom and thus to explore potential practical ways in which organ and tissue doses may be estimated when full-scale simulation is not possible in the clinic.
Izvorni jezik
Engleski
Znanstvena područja
Fizika, Kemija, Biologija
Napomena
Rad je prezentiran na skupu Workshop: Dosimetry for Second Cancer Risk Estimation EURADOS Annual Meeting Vienna 2012.
POVEZANOST RADA
Projekti:
098-0982904-2954 - Fizičko-kemijski učinci ionizirajućih zračenja u materijalima (Mihaljević, Branka, MZOS ) ( CroRIS)
Ustanove:
Institut "Ruđer Bošković", Zagreb
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus