Neutron Detection Around High Energy Medical Electron Linear Accelerators
Neutron Detection Around High Energy Medical Electron Linear Accelerators // Book of Abstracts / Goncalves, Odair Dias (ur.).
Rio de Janeiro: International Radiation Physics Society, 2012. str. 112-112 (poster, međunarodna recenzija, sažetak, znanstveni)
CROSBI ID: 621106 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Neutron Detection Around High Energy Medical Electron Linear Accelerators
Autori
Poje, Marina ; Vuković, Branko ; Radolić, Vanja ; Miklavčić, Igor ; Krpan, Ivana ; Faj, Dario ; Jurković, Slaven ; Žauhar, Gordana ; Kaliman, Zoran ; Ivković, Ana ; Planinić, Josip
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
Book of Abstracts
/ Goncalves, Odair Dias - Rio de Janeiro : International Radiation Physics Society, 2012, 112-112
Skup
12th International Symposium on Radiation Physics ISRP12
Mjesto i datum
Rio de Janeiro, Brazil, 07.10.2012. - 12.10.2012
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
radiotherapy ; electron linear accelerator ; neutron equivalent dose
Sažetak
High energy electron linear accelerators have a wide use in radiation therapy. When photon beams with photon energy higher than 10 MeV are used, unwanted neutrons can be produced in giant dipole resonance reaction between photon beam and accelerator’s head material (led, tungsten). In this reaction an isotropic flux of neutrons is produced, dominated by neutrons with energies between 700 keV and 1 MeV. Characterisation of full radiation field is important to correctly evaluate exposure of patients and medical personnel. The detection of neutrons was made around 15 MV photon beams of Siemens Mevatron MD2 and Siemens Oncor Impression as well as around 18 MV photon beams of 2 Siemens Oncor Expression linear accelerators. Measurements were performed by nuclear track etched detectors CR-39 associated with converter (radiator) that consist of 10B and with an active neutron detector Thermo BIOREM FHT 742. Detectors were set at several locations to evaluate neutron ambient dose equivalent and/or neutron dose rate to which either medical personnel or patients could be exposed. Also, neutron dose dependence on collimator aperture was analysed. The rooms of two linear accelerators, (with 15MV and 18 MV high energy beams), were built in the place of decommissioned 60Co units, so the main limitation was the space. Therefore, the mazes are very short and the plates of steel in combination with concrete were used for shielding, making the neutron flux more pronounced. In both rooms built intentionally for accelerators, the neutron exposure to the personnel was significantly lower. Measured values at the beam isocenters were comparable with the published ones, showing from 2 to 4 times higher neutron dose equivalent in 18 MV beams than in 15 MV beams. Neutron fluence increased with field size having the maximum at 20 cm x 20 cm, and then decreased with wider collimator apertures.
Izvorni jezik
Engleski
Znanstvena područja
Fizika
POVEZANOST RADA
Projekti:
MZOS-286-0000000-3212 - Radioaktivnost u okolišu, detekcija i primjena (Planinić, Josip, MZOS ) ( CroRIS)
Ustanove:
Sveučilište u Osijeku - Odjel za fiziku
Profili:
Vanja Radolić
(autor)
Gordana Žauhar
(autor)
Josip Planinić
(autor)
Zoran Kaliman
(autor)
Slaven Jurković
(autor)
Igor Miklavčić
(autor)
Marina Poje Sovilj
(autor)
Dario Faj
(autor)
Branko Vukovic
(autor)
Ivana Krpan
(autor)