Pregled bibliografske jedinice broj: 550041
Kalibracija brahiterapijskog izvora 192Ir velike brzine doze: analiza rezultata i nepouzdanosti
Kalibracija brahiterapijskog izvora 192Ir velike brzine doze: analiza rezultata i nepouzdanosti // Zbornik radova VIII.Simpozija Hrvatskog društva za zaštitu od zračenja / Krajcar Bronić, Ines ; Kopjar, Nevenka, Milić, Mirta ; Branica, Gina (ur.).
Zagreb: Hrvatsko društvo za zaštitu od zračenja, 2011. str. 113-119 (predavanje, domaća recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 550041 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Kalibracija brahiterapijskog izvora 192Ir velike brzine doze: analiza rezultata i nepouzdanosti
(High dose rate 192Ir source calibration: analysis of results and uncertainties)
Autori
Bokuluć, Tomislav ; Mrčela, Iva ; Budanec, Mirajana ; Vekić, Branko ; Fröbe, Ana ; Kusić, Zvonko
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
Zbornik radova VIII.Simpozija Hrvatskog društva za zaštitu od zračenja
/ Krajcar Bronić, Ines ; Kopjar, Nevenka, Milić, Mirta ; Branica, Gina - Zagreb : Hrvatsko društvo za zaštitu od zračenja, 2011, 113-119
ISBN
978-953-96133-7-0
Skup
Osmi simpozij Hrvatskog društav za zaštitu od zračenja s međunarodnim sudjelovanjem
Mjesto i datum
Krk, Hrvatska, 13.04.2011. - 15.04.2011
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Domaća recenzija
Ključne riječi
HDR brahiterapija; 192Ir; kalibracija
(High dose rate brachytherapy; 192 Ir; calibration)
Sažetak
A lack of primary standards for HDR 192Ir brachytherapy has been known for years to cause the largest problem in obtaining lower uncertainties in RAKR measurements. The currently used calibration techniques where chambers are calibrated against interim standards based on interpolation techniques, provide typically an uncertainty of ± 2.5% (k=2) in the calibration factor. It constitues the largest contribution to the total uncertainty in RAKR and may be significantly reduced through the on-going developement of primary standards. In this work, RAKR values of 23HDR 192Ir sources were measured in the department in the past 10 years of clinical HDRBT application. All differences between the measured RAKR and manufacturers specification were lower than ± 3.5%. Upon the acceptance of a more rigorous protocol, through which the different variables, such as thermal stability and geometrical reproducibility, have been better controlled during the measurement, these differences dropped to less than ±2%. Two ionisation chambers tested in thos work showed very good precision, long term stability, small polarity effect and recombination equal to the manufacturers specification. Differences in repeated measurements of RAKR during the 6 weeks period, in which a source was also changed, were minimal ; RAKR-s stayed within 1%. The calibration technique on the re-entrant well type chamber is due to its simplicity, speed, accuracy and long term stability a highly recommended option for an institution performing HDR brachytherapy