Dosimetric verification of compensated beams using radiographic film
Dosimetric verification of compensated beams using radiographic film // Radiology and oncology, 45 (2011), 4; 310-314 doi:10.2478/v10019-011-0020-9 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 517286 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Dosimetric verification of compensated beams using radiographic film
Autori
Jurković, Slaven ; Žauhar, Gordana ; Faj, Dario ; Radojčić Smilović, Djeni ; Švabić, Manda ; Kasabašić, Mladen ; Diklić, Ana
Izvornik
Radiology and oncology (1318-2099) 45
(2011), 4;
310-314
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
radiotherapy ; dosimetry ; compensators
Sažetak
External photon beam modulation using compensators in order to achieve a desired dose distribution when brachytherapy treatment is followed by external beam radiation is a well-established technique. A compensator modulates the central part of the beam, and the dose beneath the thickest part of the compensator is delivered mostly by scattered, low energy photons. A two-dimensional detector with a good spatial resolution is needed for the verification of those beams. In this work, the influence of different types of detectors on the measured modulated dose distributions was examined. Dosimetric verification was performed using X-Omat V, Eastman Kodak radiographic films at different depths in a solid water phantom. The film measurements were compared with those made by ionization chambers. Photon beams were also modelled using EGSnrc Monte Carlo algorithm to explain the measured results. Monte Carlo calculated over-response of the film under the thickest part of the compensator was over 15%, which was confirmed by measurements. The magnitude of over-response could be associated with changes in the spectra of photon energy in the beam. The radiographic film can be used for the dosimetry of compensated high energy photon beams, with limitations in volumes where photon spectra are hardly degraded.
Izvorni jezik
Engleski
Znanstvena područja
Fizika, Temeljne medicinske znanosti
POVEZANOST RADA
Projekti:
219-2192382-2386 - Razvoj modela funkcije cirkulacijskog sustava ovisnog o arterijskoj elastičnosti (Marić, Svjetlana, MZOS ) ( CroRIS)
286-0000000-3212 - Radioaktivnost u okolišu, detekcija i primjena (Planinić, Josip, MZOS ) ( CroRIS)
316-0000000-3192 - Analiza složenih sustava spektroskopskim metodama (Orlić, Nada, MZOS ) ( CroRIS)
Ustanove:
Medicinski fakultet, Rijeka,
Klinički bolnički centar Osijek,
Medicinski fakultet, Osijek,
Klinički bolnički centar Rijeka,
Sveučilište u Osijeku - Odjel za fiziku,
Sveučilište u Rijeci - Odjel za fiziku
Profili:
Gordana Žauhar (autor)
Ana Božanić (autor)
Slaven Jurković (autor)
Manda Švabić Kolacio (autor)
Dario Faj (autor)
Mladen Kasabašić (autor)
Citiraj ovu publikaciju:
Časopis indeksira:
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE