Image quality and dose in mammography in 17 countries in Africa, Asia and Eastern Europe: Results from IAEA projects


Ciraj-Bjelac, Olivera; Avramova-Cholakova, Simona; Beganović, Adnan; Economides, Sotirios; Faj, Dario; Gershan, Vesna; Grupetta, Edward; Kharitah, M.H.; Milaković, Milomir; Milu, Constantin et al.
Image quality and dose in mammography in 17 countries in Africa, Asia and Eastern Europe: Results from IAEA projects // European journal of radiology, 81 (2012), 9; 2161-2168 doi:10.1016/j.ejrad.2011.05.026 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 516119 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Image quality and dose in mammography in 17 countries in Africa, Asia and Eastern Europe: Results from IAEA projects

Autori
Ciraj-Bjelac, Olivera ; Avramova-Cholakova, Simona ; Beganović, Adnan ; Economides, Sotirios ; Faj, Dario ; Gershan, Vesna ; Grupetta, Edward ; Kharitah, M.H. ; Milaković, Milomir ; Milu, Constantin ; Muhogorak, W.E. ; Muthuvelu, Pirunthavany ; Oola, Samuel ; Setayeshin, Saeid ; Schandorfo, Cyril ; Ursuleanp, Ion ; Videnović, Ivan ; Zaman, Areesha ; Ziliukas, Julius ; Rehani, M.M.

Izvornik
European journal of radiology (0720-048X) 81 (2012), 9; 2161-2168

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
mammography; image quality; radiation dose; QA

Sažetak
The objective is to study mammography practice from an optimisation point of view by assessing the impact of simple and immediately implementable corrective actions on image quality. This prospective multinational study included 54 mammography units in 17 countries. More than 21, 000 mammography images were evaluated using a three-level image quality scoring system. Following initial assessment, appropriate corrective actions were implemented and image quality was re-assessed in 24 units. The fraction of images that were considered acceptable without any remark in the first phase (before the implementation of corrective actions) was 70% and 75% for cranio-caudal and medio-lateral oblique projections, respectively. The main causes for poor image quality before corrective actions were related to film processing, damaged or scratched image receptors, or film-screen combinations that are not spectrally matched, inappropriate radiographic techniques and lack of training. Average glandular dose to a standard breast was 1.5 mGy (mean and range 0.59–3.2 mGy). After optimisation the frequency of poor quality images decreased, but the relative contributions of the various causes remained similar. Image quality improvements following appropriate corrective actions were up to 50 percentage points in some facilities. Poor image quality is a major source of unnecessary radiation dose to the breast. An increased awareness of good quality mammograms is of particular importance for countries that are moving towards introduction of population-based screening programmes. The study demonstrated how simple and low-cost measures can be a valuable tool in improving of image quality in mammography.

Izvorni jezik
Engleski

Znanstvena područja
Fizika, Kliničke medicinske znanosti



POVEZANOST RADA


Projekti:
219-2192382-2386 - Razvoj modela funkcije cirkulacijskog sustava ovisnog o arterijskoj elastičnosti (Marić, Svjetlana, MZOS ) ( CroRIS)
286-0000000-3212 - Radioaktivnost u okolišu, detekcija i primjena (Planinić, Josip, MZOS ) ( CroRIS)

Ustanove:
Klinički bolnički centar Osijek,
Medicinski fakultet, Osijek,
Sveučilište u Osijeku - Odjel za fiziku

Profili:

Avatar Url Dario Faj (autor)

Poveznice na cjeloviti tekst rada:

doi www.sciencedirect.com ac.els-cdn.com dx.doi.org

Citiraj ovu publikaciju:

Ciraj-Bjelac, Olivera; Avramova-Cholakova, Simona; Beganović, Adnan; Economides, Sotirios; Faj, Dario; Gershan, Vesna; Grupetta, Edward; Kharitah, M.H.; Milaković, Milomir; Milu, Constantin et al.
Image quality and dose in mammography in 17 countries in Africa, Asia and Eastern Europe: Results from IAEA projects // European journal of radiology, 81 (2012), 9; 2161-2168 doi:10.1016/j.ejrad.2011.05.026 (međunarodna recenzija, članak, znanstveni)
Ciraj-Bjelac, O., Avramova-Cholakova, S., Beganović, A., Economides, S., Faj, D., Gershan, V., Grupetta, E., Kharitah, M., Milaković, M. & Milu, C. (2012) Image quality and dose in mammography in 17 countries in Africa, Asia and Eastern Europe: Results from IAEA projects. European journal of radiology, 81 (9), 2161-2168 doi:10.1016/j.ejrad.2011.05.026.
@article{article, author = {Ciraj-Bjelac, Olivera and Avramova-Cholakova, Simona and Beganovi\'{c}, Adnan and Economides, Sotirios and Faj, Dario and Gershan, Vesna and Grupetta, Edward and Kharitah, M.H. and Milakovi\'{c}, Milomir and Milu, Constantin and Muhogorak, W.E. and Muthuvelu, Pirunthavany and Oola, Samuel and Setayeshin, Saeid and Schandorfo, Cyril and Ursuleanp, Ion and Videnovi\'{c}, Ivan and Zaman, Areesha and Ziliukas, Julius and Rehani, M.M.}, year = {2012}, pages = {2161-2168}, DOI = {10.1016/j.ejrad.2011.05.026}, keywords = {mammography, image quality, radiation dose, QA}, journal = {European journal of radiology}, doi = {10.1016/j.ejrad.2011.05.026}, volume = {81}, number = {9}, issn = {0720-048X}, title = {Image quality and dose in mammography in 17 countries in Africa, Asia and Eastern Europe: Results from IAEA projects}, keyword = {mammography, image quality, radiation dose, QA} }
@article{article, author = {Ciraj-Bjelac, Olivera and Avramova-Cholakova, Simona and Beganovi\'{c}, Adnan and Economides, Sotirios and Faj, Dario and Gershan, Vesna and Grupetta, Edward and Kharitah, M.H. and Milakovi\'{c}, Milomir and Milu, Constantin and Muhogorak, W.E. and Muthuvelu, Pirunthavany and Oola, Samuel and Setayeshin, Saeid and Schandorfo, Cyril and Ursuleanp, Ion and Videnovi\'{c}, Ivan and Zaman, Areesha and Ziliukas, Julius and Rehani, M.M.}, year = {2012}, pages = {2161-2168}, DOI = {10.1016/j.ejrad.2011.05.026}, keywords = {mammography, image quality, radiation dose, QA}, journal = {European journal of radiology}, doi = {10.1016/j.ejrad.2011.05.026}, volume = {81}, number = {9}, issn = {0720-048X}, title = {Image quality and dose in mammography in 17 countries in Africa, Asia and Eastern Europe: Results from IAEA projects}, keyword = {mammography, image quality, radiation dose, QA} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font