Evaluation of the Innovance D-DIMER analytical performance
Evaluation of the Innovance D-DIMER analytical performance // Clinical chemistry and laboratory medicine, 47 (2009), 8; 945-951 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 478843 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Evaluation of the Innovance D-DIMER analytical performance
Autori
Coen Herak, Desiree ; Miloš, Marija ; Zadro, Renata
Izvornik
Clinical chemistry and laboratory medicine (1434-6621) 47
(2009), 8;
945-951
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
coagulation; D-dimer; evaluation; particleenhanced
Sažetak
Background: Widespread use of D-dimer in recent years has led to the development of a number of new fully automated quantitative D-dimer assays. Methods: We evaluated the analytical performance of the particle-enhanced immunoturbidimetric assay Innovance D-DIMER (Siemens Medical Solutions) on the Behring Coagulation System (BCS) analyzer. Results: Within-run coefficients of variation (CVs) for samples with low, borderline, slightly, and extremely increased D-dimer concentrations were 2.1%–5.5%, whereas between-run CVs for control samples with low and extremely increased D-dimer were 5.5%– 8.4%. The assay exhibited good linearity in the working range between 0.17 mg/L and 5.45 mg/L fibrinogen equivalent units (FEU), with the lower limit of detection of 0.099 mg/L FEU. The upper reference value determined in 40 plasma samples from healthy volunteers was 0.495 mg/L FEU. The results obtained in 457 fresh plasma samples were compared with results obtained with VIDAS D-Dimer Exclusion. Passing and Bablok regression analysis demonstrated highly significant correlation (ys1.370x–0.108, rs0.952, p-0.001). Bland and Altman difference plots demonstrated slightly higher results obtained with Innovance D-DIMER that was more pronounced with increasing values. Very good agreement between both assays was observed (ks0.860 ; 95% confidence interval (CI), 0.811–0.908). Conclusions: This study demonstrates that Innovance D-DIMER fulfills all analytical requirements for daily routine use.
Izvorni jezik
Engleski
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE
Uključenost u ostale bibliografske baze podataka::
- MEDLINE