Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 443908

Overview of RPL investigations at the Ruđer Bošković Institute


Knežević, Željka; Miljanić, Saveta; Milković, Đurđica; Ranogajec-Komor, Maria
Overview of RPL investigations at the Ruđer Bošković Institute // Proceedings of the Fifth International Workshop on Individual Monitoring of Ionizing Radiation / Yamamoto, T. (ur.).
Ōarai: Chiyoda Technol Corporation (CTC), 2010. str. 123-131 (pozvano predavanje, nije recenziran, pp prezentacija, znanstveni)


CROSBI ID: 443908 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Overview of RPL investigations at the Ruđer Bošković Institute

Autori
Knežević, Željka ; Miljanić, Saveta ; Milković, Đurđica ; Ranogajec-Komor, Maria

Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, pp prezentacija, znanstveni

Izvornik
Proceedings of the Fifth International Workshop on Individual Monitoring of Ionizing Radiation / Yamamoto, T. - Ōarai : Chiyoda Technol Corporation (CTC), 2010, 123-131

Skup
Fifth International Workshop on Individual Monitoring of Ionizing Radiation

Mjesto i datum
Ōarai, Japan, 28.11.2009. - 29.11.2009

Vrsta sudjelovanja
Pozvano predavanje

Vrsta recenzije
Nije recenziran

Ključne riječi
Radiophotoluminescent (RPL) dosimetry

Sažetak
Radiophotoluminescent (RPL) glass dosimetry has been introduced in Radiation Chemistry and Dosimetry Laboratory (RCDL) in year 2005. Thanks to the collaboration between RCDL and Chiyoda Technol Corporation the laboratory received two radiophotoluminescent readers: FGD-202 RPL reader and SC-1 dosemeters are primarily for the measurements in the field of environmental dosimetry. Dose Ace FGD -1000 reader and small glass rod dosemetrs are used for investigations in the field of medical dosimetry. The characterisation of RPL dosemeters according to international standards IEC 1066 (1991) and IEC 61066 (2006) was carried out. Dosimetric characteristics such as reproducibility, linearity, detection threshold and energy dependence for both dosimetry systems were determined. The results of investigated SC-1 RPL dosemeters showed that they fulfil the requirements of international standards for environmental dosimetry. Recently they are together with TLDs included in the field measurements around the nuclear power plant Krško. Results received in 3 measurement cycles of 6 months will be presented. In the cooperation with Children’s Hospital Srebrnjak doses on children during diagnostic X-ray procedures were measured with both RPL dosimetry systems. The measurements of the out-of-field doses during the Leksell Gamma Knife Model C radiosurgery for children were started. The children population was chosen due to their higher susceptibility to radiation. Preliminary results obtained for two children show very good agreement between different dosimetry methods. In the future the extension of RPL measurements in medicine during various radiology and radiotherapy procedures and the comparison of calculated doses with dosimetry measurements in RANDO phantom are planned.

Izvorni jezik
Engleski

Znanstvena područja
Fizika, Kemija



POVEZANOST RADA


Projekti:
098-0982904-2954 - Fizičko-kemijski učinci ionizirajućih zračenja u materijalima (Mihaljević, Branka, MZOS ) ( CroRIS)

Ustanove:
Institut "Ruđer Bošković", Zagreb


Citiraj ovu publikaciju:

Knežević, Željka; Miljanić, Saveta; Milković, Đurđica; Ranogajec-Komor, Maria
Overview of RPL investigations at the Ruđer Bošković Institute // Proceedings of the Fifth International Workshop on Individual Monitoring of Ionizing Radiation / Yamamoto, T. (ur.).
Ōarai: Chiyoda Technol Corporation (CTC), 2010. str. 123-131 (pozvano predavanje, nije recenziran, pp prezentacija, znanstveni)
Knežević, Ž., Miljanić, S., Milković, Đ. & Ranogajec-Komor, M. (2010) Overview of RPL investigations at the Ruđer Bošković Institute. U: Yamamoto, T. (ur.)Proceedings of the Fifth International Workshop on Individual Monitoring of Ionizing Radiation.
@article{article, author = {Kne\v{z}evi\'{c}, \v{Z}eljka and Miljani\'{c}, Saveta and Milkovi\'{c}, \DJur\djica and Ranogajec-Komor, Maria}, editor = {Yamamoto, T.}, year = {2010}, pages = {123-131}, keywords = {Radiophotoluminescent (RPL) dosimetry}, title = {Overview of RPL investigations at the Ru\djer Bo\v{s}kovi\'{c} Institute}, keyword = {Radiophotoluminescent (RPL) dosimetry}, publisher = {Chiyoda Technol Corporation (CTC)}, publisherplace = {\={O}arai, Japan} }
@article{article, author = {Kne\v{z}evi\'{c}, \v{Z}eljka and Miljani\'{c}, Saveta and Milkovi\'{c}, \DJur\djica and Ranogajec-Komor, Maria}, editor = {Yamamoto, T.}, year = {2010}, pages = {123-131}, keywords = {Radiophotoluminescent (RPL) dosimetry}, title = {Overview of RPL investigations at the Ru\djer Bo\v{s}kovi\'{c} Institute}, keyword = {Radiophotoluminescent (RPL) dosimetry}, publisher = {Chiyoda Technol Corporation (CTC)}, publisherplace = {\={O}arai, Japan} }




Contrast
Increase Font
Decrease Font
Dyslexic Font