Dangerous High Frequency Current Leakage During Neurophysiological Intraoperative Monitoring
Dangerous High Frequency Current Leakage During Neurophysiological Intraoperative Monitoring // Symposium on Intraoperative Neurophysiology with the 19th Dr. Janez Faganel memorial Lecture / Rodi, Zoran ; Deletis, Vedran (ur.).
Ljubljana: Institute of Clinical Neurophysiology, Division of Neurology, University Medical Center, Ljubljana, 2003. str. 65-68 (predavanje, međunarodna recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 302946 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Dangerous High Frequency Current Leakage During Neurophysiological Intraoperative Monitoring
Autori
Išgum, Velimir ; Deletes, Vedran
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
Symposium on Intraoperative Neurophysiology with the 19th Dr. Janez Faganel memorial Lecture
/ Rodi, Zoran ; Deletis, Vedran - Ljubljana : Institute of Clinical Neurophysiology, Division of Neurology, University Medical Center, Ljubljana, 2003, 65-68
ISBN
961-90106-5-5
Skup
Symposium on Intraoperative Neurophysiology with the 19th Dr. Janez Faganel memorial Lecture
Mjesto i datum
Ljubljana, Slovenija, 17.11.2003. - 18.11.2003
Vrsta sudjelovanja
Predavanje
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
evoked potential monitoring; electrical safety; high frequency leakage current
Sažetak
We describe incidents where six patients undergoing neurosurgical procedures sustained first and second degree burns under the skin where stimulating electrodes had been placed during evoked potential monitoring, following the use of monopolar, cautericing device. A post-operative test of all 18 monopolar, cauterising devices used in the operating theatres reveald one (1) with a high frequency (HF) current leakage of 287 mA (when in ‘ coagulation pin point’ mode) and 442 mA (when in ‘ coagulation spray’ mode). This was 2.5 to 3 times above safety limits. We propose a solution for preventing these type of incidents.