Comparison of different dosimetry system for dose measurements in diagnostic radiology
Comparison of different dosimetry system for dose measurements in diagnostic radiology // Radiation Protection: from Knowledge to Action, Proceedings of Second European IRPA Congress
Pariz: Société Françarse de Radioprotection, 2006. str. P-091 (poster, nije recenziran, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 270670 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Comparison of different dosimetry system for dose measurements in diagnostic radiology
Autori
Milković, Đurđica ; Ranogajec-Komor, Mária ; Miljanić, Saveta ; Knežević, Željka ; Krpan, Katarina
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
Radiation Protection: from Knowledge to Action, Proceedings of Second European IRPA Congress
/ - Pariz : Société Françarse de Radioprotection, 2006, P-091
Skup
Second European IRPA Congress
Mjesto i datum
Pariz, Francuska, 15.05.2006. - 19.05.2006
Vrsta sudjelovanja
Poster
Vrsta recenzije
Nije recenziran
Ključne riječi
TLD; RPL dosimetry; X-ray diagnostic
Sažetak
The dose measurement on patients in X-ray diagnostic is not simple, because low doses with low and various energies have to be measured. The aim of this preliminary study was to compare high sensitivity thermoluminescent dosimeter (TLD) (LiF:Mg, Cu, P) and radiophotoluminescent (RPL) glass dosimeters for dose measurements in routine X-ray diagnostic of chest of children. The energy dependence of the dosimeters was investigated in Secondary Standard Dosimetry Laboratory (SSDL). The energy range was 33-65 keV mean energy, the dosimeteres were placed “ free in air” and on the water phantom. The results were compared to calculated values of Hp(10). The next step was the irradiation in a routine X-ray diagnostic unit. Irradiations were performed by the Shimadzu X-ray unit. The selected irradiation conditions were the same as that most commonly used for baby examinations. Doses were measured with dosimeters placed “ free-in-air” and also with the dosimeters placed on the water phantom and “ baby phantom” . The results show that the RPL glass dosimeters and LiF:Mg, Cu, P based TLDs are suitable for low dose measurements in X-ray diagnostic. The uncertainty of dose determination is mainly caused by the energy dependence of dosimeters.
Izvorni jezik
Engleski
Znanstvena područja
Kemija
POVEZANOST RADA
Projekti:
0098063
277-2770966-0962 - Učinci radioloških dijagnostičkih postupaka pri pregledu dišnih putova u djece (Milković, Đurđica, MZOS ) ( CroRIS)
Ustanove:
Institut "Ruđer Bošković", Zagreb
Profili:
Željka Knežević Medija
(autor)
Saveta Miljanić
(autor)
Ferencne Ranogajec
(autor)
Đurđica Milković
(autor)