Pregled bibliografske jedinice broj: 258540
The construction of compensators for MV photon beams
The construction of compensators for MV photon beams // 2nd Austrian, Italian, Slovenian and Croatian Medical Physics Meeting : proceedings / Kovačević, Nenad (ur.).
Zagreb: Faculty of Electrical Engineering and Computing-ZESOI ; Medical Physics Division of Croatian Medical and Biological Engineering Society, 2006. (poster, domaća recenzija, cjeloviti rad (in extenso), znanstveni)
CROSBI ID: 258540 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
The construction of compensators for MV photon beams
Autori
Jurković, Slaven ; Bistrović, Matija ; Faj, Dario ; Žauhar, Gordana ; Matić, Zoran ; Smilović Radojčić, Đeni
Vrsta, podvrsta i kategorija rada
Radovi u zbornicima skupova, cjeloviti rad (in extenso), znanstveni
Izvornik
2nd Austrian, Italian, Slovenian and Croatian Medical Physics Meeting : proceedings
/ Kovačević, Nenad - Zagreb : Faculty of Electrical Engineering and Computing-ZESOI ; Medical Physics Division of Croatian Medical and Biological Engineering Society, 2006
Skup
Austrian, Italian, Slovenian and Croatian Medical Physics Meeting (2 ; 2006)
Mjesto i datum
Opatija, Hrvatska, 28.04.2006. - 29.04.2006
Vrsta sudjelovanja
Poster
Vrsta recenzije
Domaća recenzija
Ključne riječi
radiation therapy; compensators
Sažetak
The use of compensators in order to achieve desired dose distribution had a long history and is well-established technique in radiation therapy planning. There are several different methods of calculation for compensator’ s shape discussed in literature. An alternative method is presented. The method is based on Cunningham's modification of Clarkson's method to calculate scattered radiation in beams with inhomogeneous cross-sectional dose distribution. The algorithm is rather general and could be applied to different required dose distributions in various clinical situations, different photon beams, and various compensating materials. The algorithm is feasible, quickly computable and versatile. The method is verified using film dosimetry in several practical situations manufacturing compensators out of steel granules and low melting alloy. Additional measurements were made using electronic portal imaging developed in our hospital. The results are compared with effective attenuation method.
Izvorni jezik
Engleski
Znanstvena područja
Fizika
POVEZANOST RADA
Projekti:
TP-02/0062-4
Ustanove:
Medicinski fakultet, Rijeka,
Klinika za tumore,
Klinički bolnički centar Osijek,
Klinički bolnički centar Rijeka
Profili:
Matija Bistrović
(autor)
Slaven Jurković
(autor)
Dario Faj
(autor)
Đeni Smilović Radojčić
(autor)
Gordana Žauhar
(autor)