Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 258274

Comparison of enzymatic assay for HbA1c measurement (Abbott Architect) with capillary electrophoresis (Sebia Minicap Flex piercing analyser)


Tešija Kuna, Andrea; Dukić, Kristina; Nikolac Gabaj, Nora; Miler, Marijana; Vukasović, Ines; Langer, Sanja; Šimundić, Ana-Maria; Vrkić, Nada
Comparison of enzymatic assay for HbA1c measurement (Abbott Architect) with capillary electrophoresis (Sebia Minicap Flex piercing analyser) // Laboratory medicine, 49 (2018), 3; 231-238 doi:10.1093/labmed/lmx090 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 258274 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Comparison of enzymatic assay for HbA1c measurement (Abbott Architect) with capillary electrophoresis (Sebia Minicap Flex piercing analyser)

Autori
Tešija Kuna, Andrea ; Dukić, Kristina ; Nikolac Gabaj, Nora ; Miler, Marijana ; Vukasović, Ines ; Langer, Sanja ; Šimundić, Ana-Maria ; Vrkić, Nada

Izvornik
Laboratory medicine (0007-5027) 49 (2018), 3; 231-238

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
glycated hemoglobin ; electrophoresis ; enzymatic methods ; imprecison ; lipemia ; fetal hemoglobin

Sažetak
Objective: The HbA1c assays are required to have high analytical precision and low variability between different assays. We aimed to compare the analytical performances of enzymatic method and capillary electrophoresis (CE) for HbA1c measurement. Methods: Imprecision, carryover, stability, linearity, method comparison and interferences were evaluated for HbA1c enzymatic method (Abbott, USA) and CE (Sebia, France). Results: Both methods have shown overall within-laboratory imprecision <3% for IFCC units (<2% NGSP units), with %CV <2 for medium and high HbA1c concentration. Carryover effects were within acceptable criteria. The linearity of both methods has proven to be excellent (R2 = 0.999). Significant proportional and constant difference were found for enzymatic method in comparison to CE, but not clinically relevant (<5 mmol/mol ; NGSP < 0.5%). At the clinically relevant HbA1c concentration stability observed with CE and enzymatic method were acceptable (bias <3%). Enzymatic method showed to be subjected to interference of triglyceride 8.11 mmol/L and CE to interference of HbF 10.6%. Conclusion: Enzymatic method proved to be comparable to CE method in analytical performances but one should be aware that certain interferences can influence the measurements of each method.

Izvorni jezik
Engleski

Znanstvena područja
Kliničke medicinske znanosti, Farmacija



POVEZANOST RADA


Ustanove:
Farmaceutsko-biokemijski fakultet, Zagreb,
Klinička bolnica "Sveti Duh",
KBC "Sestre Milosrdnice"

Poveznice na cjeloviti tekst rada:

doi academic.oup.com

Citiraj ovu publikaciju:

Tešija Kuna, Andrea; Dukić, Kristina; Nikolac Gabaj, Nora; Miler, Marijana; Vukasović, Ines; Langer, Sanja; Šimundić, Ana-Maria; Vrkić, Nada
Comparison of enzymatic assay for HbA1c measurement (Abbott Architect) with capillary electrophoresis (Sebia Minicap Flex piercing analyser) // Laboratory medicine, 49 (2018), 3; 231-238 doi:10.1093/labmed/lmx090 (međunarodna recenzija, članak, znanstveni)
Tešija Kuna, A., Dukić, K., Nikolac Gabaj, N., Miler, M., Vukasović, I., Langer, S., Šimundić, A. & Vrkić, N. (2018) Comparison of enzymatic assay for HbA1c measurement (Abbott Architect) with capillary electrophoresis (Sebia Minicap Flex piercing analyser). Laboratory medicine, 49 (3), 231-238 doi:10.1093/labmed/lmx090.
@article{article, author = {Te\v{s}ija Kuna, Andrea and Duki\'{c}, Kristina and Nikolac Gabaj, Nora and Miler, Marijana and Vukasovi\'{c}, Ines and Langer, Sanja and \v{S}imundi\'{c}, Ana-Maria and Vrki\'{c}, Nada}, year = {2018}, pages = {231-238}, DOI = {10.1093/labmed/lmx090}, keywords = {glycated hemoglobin, electrophoresis, enzymatic methods, imprecison, lipemia, fetal hemoglobin}, journal = {Laboratory medicine}, doi = {10.1093/labmed/lmx090}, volume = {49}, number = {3}, issn = {0007-5027}, title = {Comparison of enzymatic assay for HbA1c measurement (Abbott Architect) with capillary electrophoresis (Sebia Minicap Flex piercing analyser)}, keyword = {glycated hemoglobin, electrophoresis, enzymatic methods, imprecison, lipemia, fetal hemoglobin} }
@article{article, author = {Te\v{s}ija Kuna, Andrea and Duki\'{c}, Kristina and Nikolac Gabaj, Nora and Miler, Marijana and Vukasovi\'{c}, Ines and Langer, Sanja and \v{S}imundi\'{c}, Ana-Maria and Vrki\'{c}, Nada}, year = {2018}, pages = {231-238}, DOI = {10.1093/labmed/lmx090}, keywords = {glycated hemoglobin, electrophoresis, enzymatic methods, imprecison, lipemia, fetal hemoglobin}, journal = {Laboratory medicine}, doi = {10.1093/labmed/lmx090}, volume = {49}, number = {3}, issn = {0007-5027}, title = {Comparison of enzymatic assay for HbA1c measurement (Abbott Architect) with capillary electrophoresis (Sebia Minicap Flex piercing analyser)}, keyword = {glycated hemoglobin, electrophoresis, enzymatic methods, imprecison, lipemia, fetal hemoglobin} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font