Pregled bibliografske jedinice broj: 258012
Intercomparison of dosimetry systems based on CaF_2:Mn TL detectors
Intercomparison of dosimetry systems based on CaF_2:Mn TL detectors // Radiation Protection Dosimetry, 119 (2006), 1-4; 300-305 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 258012 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Intercomparison of dosimetry systems based on CaF_2:Mn TL detectors
Autori
Zorko, Benjamin ; Miljanić, Saveta ; Vekić, Branko ; Štuhec, Matjaž ; Gobec, Sandi ; Ranogajec-Komor, Maria
Izvornik
Radiation Protection Dosimetry (0144-8420) 119
(2006), 1-4;
300-305
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
TL detectors; CaF_2:Mn; intercomparison
Sažetak
The responses of readings by the TL dosimetry system MR200 TL developed in-house and used at JSI and the TOLEDO TL system used at RBI are compared. Ten measurements at different doses ranging from 0.01 mSv to 5 Sv were carried out. A set of 36 dosemeters with three pellets of CaF_2:Mn were irradiated in radiation fields of 137^Cs and 60^Co. Analysis of the measured results shows that at doses below 0.1 Sv, readers' outputs do not differ more than 5%from each other. At doses more than 1 Sv, the results obtained by the MR200 reader must be corrected with a known factor. Finally, the reproducibility of the results from the MR200 was tested.
Izvorni jezik
Engleski
Znanstvena područja
Kemija
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE