Pregled bibliografske jedinice broj: 249044
Comparison of four models for calculation of collimator scatter factors of lina cphoton beams
Comparison of four models for calculation of collimator scatter factors of lina cphoton beams // Radiology and oncology, 33 (1999), 4; 309-313 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 249044 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Comparison of four models for calculation of collimator scatter factors of lina cphoton beams
Autori
Faj, Dario ; Bistrović, Matija
Izvornik
Radiology and oncology (1318-2099) 33
(1999), 4;
309-313
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
radiotherapy; collimator scatter; photon
Sažetak
Two approaches for aproximation of collimator scatter correction factors of rectangular fields can be found in recent publications. One is based on empirical equations or some more sophisticated physical methods using certain parameters which have to be adjusted for a specific machine. The other is based on idea of decomposition of collimator scatter correction factor of two variables - into the product of two functions of one variable. In this work four models, based on decomposition are compared. The measurements were carried out using nominal 6MV and 15 MV photon beams of a Siemens Mevatron MD2 and Varian Clinac 1800 accelerators. It was shown that better approximation can be achieved with a suitable choice of basic measurements and normalisation data.
Izvorni jezik
Engleski
Znanstvena područja
Fizika, Temeljne medicinske znanosti
POVEZANOST RADA
Ustanove:
Arhitektonski fakultet, Zagreb,
Klinika za tumore,
Klinički bolnički centar Osijek
Citiraj ovu publikaciju:
Časopis indeksira:
- Scopus
Uključenost u ostale bibliografske baze podataka::
- Chemical Abstracts
- Excerpta Medica