Pregled bibliografske jedinice broj: 139745
Tolerancing and budget in design and production of test lenses for calibration of focimeters
Tolerancing and budget in design and production of test lenses for calibration of focimeters // Optometry and Vision Science, 80 (2003), 12; 839-845 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 139745 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Tolerancing and budget in design and production of test lenses for calibration of focimeters
Autori
Risović, Dubravko ; Parlić-Risović, Tijana
Izvornik
Optometry and Vision Science (1040-5488) 80
(2003), 12;
839-845
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
focimeter; test lenses; calibration; uncertainty budget; ISO 9342:1996
Sažetak
Focimeter is one of the basic clinical instruments used in every optometric practice and verification of the accuracy and calibration of the instrument are of the outmost importance. The calibration is accomplished using special test lenses. These lenses must be of high quality and of nominal ophthalmic power that is known with high accuracy. The standard ISO 9342:1996 “ Optics and Optical instruments – Test Lenses for Calibration of Focimeters” imposes stringent requirements on permitted deviation from nominal ophthalmic power of test lenses. In this study the influence of tolerancing and uncertainties in design and production of standard test lenses for calibration of focimeters is analyzed and discussed. Two avenues of approach were used: in the first the paraxial approximation was used to relate the lens parameters with ophthalmic power and to calculate the uncertainty budget. In the other approach exact ray tracing using a professional optical design program and tolerancing sensitivity analysis in conjunction with Monte Carlo simulation were used to evaluate the uncertainty budget and balance the required tolerances. The results of both approaches are compared and discussed in detail. It is shown that the customary uncertainty budgeting based on simple paraxial approximation does not guarantee the compliance with the requirements of the ISO 9342:1996 standard. Consequently, such approach proves to be inadequate for characterization and tolerancing of test lenses for calibration of focimeters.
Izvorni jezik
Engleski
Znanstvena područja
Fizika
POVEZANOST RADA
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE
Uključenost u ostale bibliografske baze podataka::
- Biological Abstracts