CONTRIBUTION OF THE AUTOMATION OF PRE-ANALYTICAL PROCESSES TO A SHORTER TURNAROUND TIME


Radeljak, Andrea; Ivić, Matea; Perkov, Sonja; Kardum Paro, Mirjana Mariana
CONTRIBUTION OF THE AUTOMATION OF PRE-ANALYTICAL PROCESSES TO A SHORTER TURNAROUND TIME // "Analytical technologies and applications" Clinical Chemistry and Laboratory Medicine (CCLM), vol. 61, no. s1, 2023, pp. s182-s323
Rim, Italija: Walter de Gruyter, 2023. str. s182-s182 doi:10.1515/cclm-2023-7036 (poster, međunarodna recenzija, sažetak, stručni)


CROSBI ID: 1278832 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
CONTRIBUTION OF THE AUTOMATION OF PRE-ANALYTICAL PROCESSES TO A SHORTER TURNAROUND TIME

Autori
Radeljak, Andrea ; Ivić, Matea ; Perkov, Sonja ; Kardum Paro, Mirjana Mariana

Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, stručni

Izvornik
"Analytical technologies and applications" Clinical Chemistry and Laboratory Medicine (CCLM), vol. 61, no. s1, 2023, pp. s182-s323 / - : Walter de Gruyter, 2023, S182-s182

Skup
WorldLab EuroMedLab Roma 2023, XXV IFCC-EFLM EuroMedLab Congress

Mjesto i datum
Rim, Italija, 21.05.2023. - 25.05.2023

Vrsta sudjelovanja
Poster

Vrsta recenzije
Međunarodna recenzija

Ključne riječi
automation of pre- analytical processes, TAT

Sažetak
BACKGROUND-AIM. Determination of vitamin D concentration by liquid chromatography-tandem mass spectrometry (LC-MS/MS) has demanding pretreatment of samples. Therefore, the overall laboratory process is time-consuming and the laboratory report is waited longer compared to immunoassays. The solutions for progressive increase in the number of requests for vitamin D determination has been found in the automated sample pretreatment. The aim of this study was to compare the turnaround time (TAT) for vitamin D in serum after the introduction of automation compared to the period when samples were prepared manually. METHODS. Automated sample pretreatment is performed on the preanalytical module Shimadzu CLAM-2030 by the modified protocol for manual preparation using commercial reagent kit (RECIPE ClinMass® LC-MS/MS 25-OH Vitamin D2/D3). Concentration of vitamin D in serum was determined by the LC-MS/MS (UPLC NEXERA X2-LCMS-8050, Shimadzu), accredited according to ISO 15189. Results for the TAT were obtained from the laboratory's information system. We analyzed the percentage of requests reported within 2 days, 3 days and 7 days. Also, we analyzed the time in which more than 99% of requests were reported. RESULTS. For the period of manual sample preparation (from January to August 2022.) requests were reported as follows: within 2 days 22, 98%, 3 days 43, 38%, 7 days 82, 6% and >99% of requests were reported within 13 days. After implementation of automated pretreatment (from September to December 2022.) requests were reported as follows: within 2 days 54, 98%, 3 days 71, 87%, 7 days 97, 54% and >99% of requests were reported within 8 days. CONCLUSIONS. A comparison of the TAT for the determination of vitamin D concentration in serum after the introduction of automated sample pretreatment compared to the period when samples were prepared manually showed that the time from receiving the request to reporting the findings was significantly reduced. This enables faster notification of clinicians and timely introduction of the necessary therapy, contributing to a more favorable treatment outcome.

Izvorni jezik
Engleski

Znanstvena područja
Kliničke medicinske znanosti



POVEZANOST RADA


Ustanove:
Klinička bolnica "Merkur"

Poveznice na cjeloviti tekst rada:

Pristup cjelovitom tekstu rada doi www.degruyter.com

Citiraj ovu publikaciju:

Radeljak, Andrea; Ivić, Matea; Perkov, Sonja; Kardum Paro, Mirjana Mariana
CONTRIBUTION OF THE AUTOMATION OF PRE-ANALYTICAL PROCESSES TO A SHORTER TURNAROUND TIME // "Analytical technologies and applications" Clinical Chemistry and Laboratory Medicine (CCLM), vol. 61, no. s1, 2023, pp. s182-s323
Rim, Italija: Walter de Gruyter, 2023. str. s182-s182 doi:10.1515/cclm-2023-7036 (poster, međunarodna recenzija, sažetak, stručni)
Radeljak, A., Ivić, M., Perkov, S. & Kardum Paro, M. (2023) CONTRIBUTION OF THE AUTOMATION OF PRE-ANALYTICAL PROCESSES TO A SHORTER TURNAROUND TIME. U: "Analytical technologies and applications" Clinical Chemistry and Laboratory Medicine (CCLM), vol. 61, no. s1, 2023, pp. s182-s323 doi:10.1515/cclm-2023-7036.
@article{article, author = {Radeljak, Andrea and Ivi\'{c}, Matea and Perkov, Sonja and Kardum Paro, Mirjana Mariana}, year = {2023}, pages = {s182-s182}, DOI = {10.1515/cclm-2023-7036}, keywords = {automation of pre- analytical processes, TAT}, doi = {10.1515/cclm-2023-7036}, title = {CONTRIBUTION OF THE AUTOMATION OF PRE-ANALYTICAL PROCESSES TO A SHORTER TURNAROUND TIME}, keyword = {automation of pre- analytical processes, TAT}, publisher = {Walter de Gruyter}, publisherplace = {Rim, Italija} }
@article{article, author = {Radeljak, Andrea and Ivi\'{c}, Matea and Perkov, Sonja and Kardum Paro, Mirjana Mariana}, year = {2023}, pages = {s182-s182}, DOI = {10.1515/cclm-2023-7036}, keywords = {automation of pre- analytical processes, TAT}, doi = {10.1515/cclm-2023-7036}, title = {CONTRIBUTION OF THE AUTOMATION OF PRE-ANALYTICAL PROCESSES TO A SHORTER TURNAROUND TIME}, keyword = {automation of pre- analytical processes, TAT}, publisher = {Walter de Gruyter}, publisherplace = {Rim, Italija} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font