Retrospective gamma analysis of patient absorbed dose distributions in a view of new tolerance criteria recommendations


Obajdin, Nevena; Smilović Radojčić, Đeni; Rajlić, David; Švabić Kolacio, Manda; Belac Lovasić, Ingrid; Jurković, Slaven
Retrospective gamma analysis of patient absorbed dose distributions in a view of new tolerance criteria recommendations // 14. hrvatski onkološki kongres
online, 2021. str. 127-128 (poster, domaća recenzija, sažetak, znanstveni)


CROSBI ID: 1254623 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Retrospective gamma analysis of patient absorbed dose distributions in a view of new tolerance criteria recommendations

Autori
Obajdin, Nevena ; Smilović Radojčić, Đeni ; Rajlić, David ; Švabić Kolacio, Manda ; Belac Lovasić, Ingrid ; Jurković, Slaven

Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni

Skup
14. hrvatski onkološki kongres

Mjesto i datum
Online, 22.04.2021. - 25.04.2021

Vrsta sudjelovanja
Poster

Vrsta recenzije
Domaća recenzija

Ključne riječi
dose calculation algorithm, pre-treatment dosimetric verification, new tolerance criteria

Sažetak
Introduction: In radiation oncology, uncertainties in treatment delivery should be kept as low as possible. Establishment of a comprehensive quality assurance (QA) program is a crucial step of uncertainties reduction. Important part of a QA program is verification of calculated patient absorbed dose distributions, also known as patient specific dosimetry (PSD). At Medical Physics Department, PSD verification is undertaken for each advanced radiation oncology treatment. Calculated absorbed dose distributions are validated using gamma method. Since 2018 international tolerance criteria for gamma passing rate calculation have become more stringent. This was a motivation for a retrospective verification of calculated dose distributions validated before 2018 according to new tolerance criteria. Methods and materials: Two different treatment planning system’s dose calculation algorithms were used for absorbed dose calculation. Standard Superposition algorithm (Elekta XiO ver. 5.10) for “field in field” (FiF) technique and Monte Carlo based algorithm (Elekta Monaco ver. 5.11) for intensity modulated radiation therapy (IMRT) technique. Detector IBA Matrixx, along with MultiCube phantom, was used for measurement of absorbed dose distributions. Linear accelerators Siemens Oncor Impression and Siemens Oncor Expression were used for dose delivery. Calculated and measured absorbed doses were compared using IBA OmniPro I’mRT. Up to 2018, recommended tolerance criteria were 3% dose-difference and 3mm distance-to-agreement (DTA). Gamma passing rate was set to ≥95%. In 2018, these criteria were changed to be 3% dose-difference and 2mm DTA. Retrospectively, absorbed dose distributions calculated using the 3%/3mm criteria were re- validated using more stringent criteria. Over 1800 FiF and over 250 IMRT dose distributions were validated. The results were presented per anatomical sites: breast, rectum, H&N, CNS, prostate, lung (FiF) and H&N, CNS, prostate (IMRT). Results and discussion: Applying the 3mm/2% criteria, for FiF, average gamma passing rates for anatomical sites, along with their standard deviation, were as follows (in %): 96.99 (1.67) for breast, 95.22 (4.12) for rectum, 96.98 (1.98) for H&N, 98.39 (2.01) for CNS, 98.17 (1.13) for prostate and 99.22 (2.32) for lung. Correspondingly, for IMRT, average gamma passing rates for anatomical sites, along with their standard deviation, were as follows (in %): 96.35 (3.95) for H&N, 97.09 (1.87) for CNS and 95.78 (1.89) for prostate. Presented results shows that despite the validation according to more stringent tolerance criteria, average gamma passing rates exceed the ≥95% tolerance limit for all observed treatment sites for both treatment planning modalities. Conclusion: Validated average gamma passing rates exceeded the ≥95% tolerance criteria when new recommendations were followed. This shows that a prerequisite to deliver prescribed absorbed dose to the patient was fulfilled and calculated absorbed dose distributions were optimized in a very good manner.

Izvorni jezik
Engleski


Citiraj ovu publikaciju:

Obajdin, Nevena; Smilović Radojčić, Đeni; Rajlić, David; Švabić Kolacio, Manda; Belac Lovasić, Ingrid; Jurković, Slaven
Retrospective gamma analysis of patient absorbed dose distributions in a view of new tolerance criteria recommendations // 14. hrvatski onkološki kongres
online, 2021. str. 127-128 (poster, domaća recenzija, sažetak, znanstveni)
Obajdin, N., Smilović Radojčić, Đ., Rajlić, D., Švabić Kolacio, M., Belac Lovasić, I. & Jurković, S. (2021) Retrospective gamma analysis of patient absorbed dose distributions in a view of new tolerance criteria recommendations. U: 14. hrvatski onkološki kongres.
@article{article, author = {Obajdin, Nevena and Smilovi\'{c} Radoj\v{c}i\'{c}, \DJeni and Rajli\'{c}, David and \v{S}vabi\'{c} Kolacio, Manda and Belac Lovasi\'{c}, Ingrid and Jurkovi\'{c}, Slaven}, year = {2021}, pages = {127-128}, keywords = {dose calculation algorithm, pre-treatment dosimetric verification, new tolerance criteria}, title = {Retrospective gamma analysis of patient absorbed dose distributions in a view of new tolerance criteria recommendations}, keyword = {dose calculation algorithm, pre-treatment dosimetric verification, new tolerance criteria}, publisherplace = {online} }
@article{article, author = {Obajdin, Nevena and Smilovi\'{c} Radoj\v{c}i\'{c}, \DJeni and Rajli\'{c}, David and \v{S}vabi\'{c} Kolacio, Manda and Belac Lovasi\'{c}, Ingrid and Jurkovi\'{c}, Slaven}, year = {2021}, pages = {127-128}, keywords = {dose calculation algorithm, pre-treatment dosimetric verification, new tolerance criteria}, title = {Retrospective gamma analysis of patient absorbed dose distributions in a view of new tolerance criteria recommendations}, keyword = {dose calculation algorithm, pre-treatment dosimetric verification, new tolerance criteria}, publisherplace = {online} }