Pregled bibliografske jedinice broj: 1247358
Rebound Tonometry over Soft Contact Lenses
Rebound Tonometry over Soft Contact Lenses // Acta Informatica Medica, 28 (2020), 3; 185-189 doi:10.5455/aim.2020.28.185-189 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 1247358 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Rebound Tonometry over Soft Contact Lenses
Autori
Krolo, Iva ; Mihaljevic, Boze ; Kasumovic, Aida ; Ravlic, Maja ; Herman, Jelena
Izvornik
Acta Informatica Medica (0353-8109) 28
(2020), 3;
185-189
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
Contact Lens, Rebound Tonometry, Icare, Hydrogel, Silicone Hydrogel.
Sažetak
Goldmann applanation tonometry (GAT) is named as a gold standard for intraocular pressure (IOP) measurement. Aim: To assess the accuracy of intraocular pressure (IOP) measurements using rebound tonometry over hydrogel and silicone hydrogel contact lenses (CLs) of different powers. Methods: This study included 117 eyes of 61 patients (12 male, 49 female), all habitual wearers of hydrogel and silicone hydrogel CLs, and none previously diagnosed with glaucoma, ocular hypertension or anterior surface disease. Five IOP measurements were taken over each eye using a rebound tonometer (Icare): with soft CLs in situ and then repeated without CLs. Lens power ranged from -9.50 to +10.00 spherical diopters and to a maximum of -0.75 cylinder diopters. Results: A significant positive correlation was found between IOP measurements with and without CLs. The difference between IOP measurements with (mean 20.74±5.19 mmHg) and without (mean 18.79±4.36 mmHg) CLs was found to be 1.95 mmHg (P <0.01). Statistical analysis was performed using the paired t-test and a correlation coefficient was calculated (r = 0.59 ; P <0.001). We have observed that increase in central corneal thickness (CCT) correlates positively with increase of measurement error of rebound tonometry (r = 0.43 ; P <0.001). C
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
Citiraj ovu publikaciju:
Časopis indeksira:
- Scopus