Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate


Čičak, Helena; Šonjić, Pavica; Šimundić, Ana-Maria
Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate // Biochemia medica, 32 (2022), 1; 010708, 8 doi:10.11613/BM.2022.010708 (međunarodna recenzija, članak, stručni)


CROSBI ID: 1218476 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate

Autori
Čičak, Helena ; Šonjić, Pavica ; Šimundić, Ana-Maria

Izvornik
Biochemia medica (1330-0962) 32 (2022), 1; 010708, 8

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, stručni

Ključne riječi
automated ESR ; Westergren method ; blood sedimentation rate ; verification ; haematology

Sažetak
Introduction: Automated erythrocyte sedimentation rate (ESR) analysers are based on different methodology than Westergren method. It is questionable whether ESR values obtained from those analysers are comparable with determined values with Westergren method. The aim was verification of the precision, method comparison and accuracy of automated ESR analysers: Roller 20PN (Alifax S.p.A., Polverara, Italy) and iSED (Alcor Scientific, Smithfield, USA). Materials and methods: Blood samples (N = 752 for Roller 20PN and N = 213 for iSED) were sampled into K2 EDTA (Kima, Italy) tubes for automated and 3.8% Na- citrate tubes (Kima, Italy) for Westergren method. The data was divided into three groups according to the ESR values obtained with the Westergren method: Group Low (L) (ESR ≤ 20 mm), Group Medium (M) (ESR 21-60 mm), and Group High (H) (ESR ≥ 61 mm). Method agreement was assessed by Bland-Altman analysis and Passing- Bablok regression. Results: Analyser iSED has shown better comparability with Westergren method (bias 0.0 (95%Cl -1.4 to 1.5) range than Roller 20 PN (bias = - 6.4 (95%Cl - 7.1 to -5.7) in the whole measuring. For Roller 20 PN, Passing-Bablok regression has shown constant and proportional difference for Groups L and M, and for iSED only for Group H. Roller 20 PN had lower sensitivity (0.51 (95%Cl: 0.45-0.57) than iSED (0.72 (95%Cl: 0.59-0.80) while they had comparable specificity (> 0.90) and accuracy (≥ 0.80) in comparison with the Westergren method. Conclusion: Both analysers are not comparable with the Westergren method and should not be used interchangeably

Izvorni jezik
Engleski

Znanstvena područja
Kliničke medicinske znanosti



POVEZANOST RADA


Ustanove:
Farmaceutsko-biokemijski fakultet, Zagreb,
Klinička bolnica "Sveti Duh"

Profili:

Avatar Url Ana-Maria Šimundić (autor)

Avatar Url Helena Čičak (autor)

Poveznice na cjeloviti tekst rada:

doi www.biochemia-medica.com

Citiraj ovu publikaciju:

Čičak, Helena; Šonjić, Pavica; Šimundić, Ana-Maria
Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate // Biochemia medica, 32 (2022), 1; 010708, 8 doi:10.11613/BM.2022.010708 (međunarodna recenzija, članak, stručni)
Čičak, H., Šonjić, P. & Šimundić, A. (2022) Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate. Biochemia medica, 32 (1), 010708, 8 doi:10.11613/BM.2022.010708.
@article{article, author = {\v{C}i\v{c}ak, Helena and \v{S}onji\'{c}, Pavica and \v{S}imundi\'{c}, Ana-Maria}, year = {2022}, pages = {8}, DOI = {10.11613/BM.2022.010708}, chapter = {010708}, keywords = {automated ESR, Westergren method, blood sedimentation rate, verification, haematology}, journal = {Biochemia medica}, doi = {10.11613/BM.2022.010708}, volume = {32}, number = {1}, issn = {1330-0962}, title = {Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate}, keyword = {automated ESR, Westergren method, blood sedimentation rate, verification, haematology}, chapternumber = {010708} }
@article{article, author = {\v{C}i\v{c}ak, Helena and \v{S}onji\'{c}, Pavica and \v{S}imundi\'{c}, Ana-Maria}, year = {2022}, pages = {8}, DOI = {10.11613/BM.2022.010708}, chapter = {010708}, keywords = {automated ESR, Westergren method, blood sedimentation rate, verification, haematology}, journal = {Biochemia medica}, doi = {10.11613/BM.2022.010708}, volume = {32}, number = {1}, issn = {1330-0962}, title = {Verification of automatic analysers Roller 20PN and iSED for measuring erythrocyte sedimentation rate}, keyword = {automated ESR, Westergren method, blood sedimentation rate, verification, haematology}, chapternumber = {010708} }

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • Science Citation Index Expanded (SCI-EXP)
    • SCI-EXP, SSCI i/ili A&HCI
  • Scopus
  • MEDLINE


Citati:





    Contrast
    Increase Font
    Decrease Font
    Dyslexic Font