Pregled bibliografske jedinice broj: 1139583
Experimental Testing and Finite Element Method Analysis of the Anterior Cruciate Ligament
Experimental Testing and Finite Element Method Analysis of the Anterior Cruciate Ligament // Medical engineering & physics, 95 (2021), 76-83 doi:10.1016/j.medengphy.2021.07.013 (međunarodna recenzija, članak, ostalo)
CROSBI ID: 1139583 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Experimental Testing and Finite Element Method
Analysis of the Anterior Cruciate Ligament
(Experimental Testing and Finite Element Method
Analysis of the Anterior Cruciate Ligament
Primary Repair with Internal Brace Augmentation)
Autori
Bonačić Bartolin, Petra ; Boixadera, Rachel ; Hudetz, Damir
Izvornik
Medical engineering & physics (1350-4533) 95
(2021);
76-83
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, ostalo
Ključne riječi
FEM analysis ; ACL tear ; ACL repair ; Experimental testing
Sažetak
The internal brace (IB) technique is a promising treatment option for repairing the proximal rupture of the anterior cruciate ligament (ACL). This paper presents a biomechanical evaluation of the IB technique. Sixteen cadaveric sheep knees underwent monotonic tensile tests, cyclic loading, and passive flexion-extension testing. Data were compared in a series of eight control specimens with an intact ACL and eight repaired specimens where the ACL was cut and repaired using the IB. In parallel with the mechanical testing, finite element analysis (FEA) was performed to investigate the influence of IB loading on the femur-ACL-tibia complex (FATC). The 3D geometry of the FATC was reconstructed from CT scans of the sheep. The IB 3D model was integrated with the 3D FATC for FEA to obtain the femur- repaired ACL with IB - tibia complex (FRA- IB- TC) group. For the intact specimens, the mean (±SD) failure load in the tensile testing was 937 N (±192 N), while for the FRA-IB-TC specimens, it was 519 N (±52 N). The FRA-IB-TC remained biomechanically stable during the cyclic loading testing. The FEA demonstrated an increase in ACL stress to 24.59 MPa and displacement values of 0.391 mm. The IB construct exhibited shear and notch effects at the button-suture-bone fixation site. Testing on this sheep model allowed us a parametric analysis of the impact of the IB repair technique. However, the results will need to be confirmed in a human model. In conclusion, although the IB technique has biomechanical drawbacks, the mechanical properties of the technique are satisfactory.
Izvorni jezik
Engleski
Znanstvena područja
Interdisciplinarne tehničke znanosti
POVEZANOST RADA
Ustanove:
Fakultet strojarstva i brodogradnje, Zagreb
Citiraj ovu publikaciju:
Časopis indeksira:
- Current Contents Connect (CCC)
- Web of Science Core Collection (WoSCC)
- Science Citation Index Expanded (SCI-EXP)
- SCI-EXP, SSCI i/ili A&HCI
- Scopus
- MEDLINE