Pregled bibliografske jedinice broj: 1111674
Characterization of passive dosimeters in proton pencil beam scanning – A EURADOS intercomparison for mailed dosimetry audits in proton therapy centres
Characterization of passive dosimeters in proton pencil beam scanning – A EURADOS intercomparison for mailed dosimetry audits in proton therapy centres // Physica medica, 82 (2021), 134-143 doi:10.1016/j.ejmp.2021.01.073 (međunarodna recenzija, članak, znanstveni)
CROSBI ID: 1111674 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Characterization of passive dosimeters in proton pencil beam scanning – A EURADOS intercomparison for mailed dosimetry audits in proton therapy centres
(Characterization of passive dosimeters in proton pencil beam scanning – A EURADOS intercomparison for mailed dosimetry audits in proton therapy)
Autori
De Saint-Hubert, M ; De Angelis, C ; Knežević, Željka ; Michalec, B ; Renierse, B ; Pyszka, E ; Stolarczyk, L ; Swakon, J ; Foltynska, G ; Wochnik, A ; Parisi, A ; Majer, Marija ; Harrison, R.M. ; Kopec, R ; Vanhavere, F ; Olko, P
Kolaboracija
EURADOS Working Group
Izvornik
Physica medica (1120-1797) 82
(2021);
134-143
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni
Ključne riječi
Alanine ; Luminescence dosimetry ; Mail auditing ; Proton therapy
(Alanine ; Luminescence dosimetry ; Mail auditing ; Proton therap)
Sažetak
The lack of mailed dosimetry audits of proton therapy centres in Europe has encouraged researchers of EURADOS Working Group 9 (WG9) to compare response of several existing passive detector systems in therapeutic pencil beam scanning. Alanine Electron Paramagnetic Resonance dosimetry systems from 3 different institutes (ISS, Italy ; UH, Belgium and IFJ PAN, Poland), natLiF:Mg, Ti (MTS-N) and natLiF:Mg, Cu, P (MCP-N) thermoluminescent do-simeters (TLDs), GD-352M radiophotoluminescent glass dosimeters (RPLGDs) and Al2O3:C optically stimulated dosimeters (OSLDs) were evaluate. Dosimeter repeatability, batch reproducibility and response in therapeutic Pencil Beam Scanning were verified for implementation as mail auditing system. Alanine detectors demonstrated the lowest linear energy transfer (LET) dependence with an agreement be-tween measured and treatment planning system (TPS) dose below 1%. The OSLDs measured on average a 6.3% lower dose compared to TPS calculation, with no significant difference between varying modulations and ranges. Both GD-352M and MCP-N measured a lower dose than the TPS and luminescent response was dependent on the LET of the therapeutic proton beam. Thermoluminescent response of MTS-N was also found to be dependent on the LET and a higher dose than TPS was measured with the most pronounced increase of 11%. As alanine detectors are characterized by the lowest energy dependence for different parameters of therapeutic pencil beam scanning they are suitable candidates for mail auditing in proton therapy. The response of luminescence detector systems have shown promises even though more careful calibration and corrections are needed for its implementation as part of a mailed dosimetry audit system.
Izvorni jezik
Engleski
Znanstvena područja
Fizika, Interdisciplinarne prirodne znanosti