Pretražite po imenu i prezimenu autora, mentora, urednika, prevoditelja

Napredna pretraga

Pregled bibliografske jedinice broj: 1095479

Optimization of paranasal sinus CT procedure: Ultra-low dose CT as a roadmap for pre-functional endoscopic sinus surgery


Diklić, Ana; Valković Zujić, Petra; Šegota, Doris; Dundara Debeljuh, Dea; Jurković, Slaven; Brambilla, Marco; Kalra, Mannudeep K.
Optimization of paranasal sinus CT procedure: Ultra-low dose CT as a roadmap for pre-functional endoscopic sinus surgery // Physica medica, 78 (2020), 195-200 doi:10.1016/j.ejmp.2020.09.014 (međunarodna recenzija, članak, znanstveni)


CROSBI ID: 1095479 Za ispravke kontaktirajte CROSBI podršku putem web obrasca

Naslov
Optimization of paranasal sinus CT procedure: Ultra-low dose CT as a roadmap for pre-functional endoscopic sinus surgery

Autori
Diklić, Ana ; Valković Zujić, Petra ; Šegota, Doris ; Dundara Debeljuh, Dea ; Jurković, Slaven ; Brambilla, Marco ; Kalra, Mannudeep K.

Izvornik
Physica medica (1120-1797) 78 (2020); 195-200

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
CT protocol, Paranasal sinuses, Radiation dose

Sažetak
Objective To assess image quality and radiation dose associated with ultra-low dose CT protocol for patients with benign paranasal sinus diseases undergoing functional endoscopic surgery (FESS). Methods We scanned the head portion of Alderson RANDO phantom on a second generation, dual-source, multidetector-row CT scanner (Siemens Definition Flash) using standard-dose and five low-dose protocols. Two radiologists assessed the image quality for each protocol to determine best ultra-low-dose protocols for imaging patients with benign paranasal sinus diseases undergoing FESS. The ultra-low-dose CT protocols were then used for scanning. Thereafter, 40 adult patients (age range 18–54 years, M:F 23:17) were scanned with the four low dose scanning protocols (10 patients per protocol). On both transverse and coronal reformatted CT images, two radiologists assessed visibility of key anatomic landmarks for FESS on a 2-point scale (1 = clear and complete visualization ; 2 = suboptimal visualization). Data were analyzed with descriptive statistics and Cohen’s kappa coefficient for interobserver agreement. Results In phantom study, the lowest dose scan protocol (CTDI vol 2.1 mGy, 70 kV, 75 mAs) was unacceptable due to poor image quality. For patient studies, both radiologists gave acceptable image quality scores for ultra-low-dose scan protocol with axial scan mode, automatic tube potential selection and tube current modulation (CTDI vol 2.2 mGy ; DLP 22.9 mGy.cm) with up to 60% lower dose compared to prior standard-dose CT (CTDI vol 5.3 mGy ; DLP 73.5 mGy.cm). Conclusions Ultra-low-dose CT protocol provides sufficient image quality for scanning patients undergoing functional endoscopic surgery for benign paranasal sinus diseases.

Izvorni jezik
Engleski

Znanstvena područja
Fizika, Temeljne medicinske znanosti



POVEZANOST RADA


Ustanove:
Medicinski fakultet, Rijeka,
Opća bolnica Pula,
Klinički bolnički centar Rijeka

Poveznice na cjeloviti tekst rada:

doi www.physicamedica.com

Citiraj ovu publikaciju:

Diklić, Ana; Valković Zujić, Petra; Šegota, Doris; Dundara Debeljuh, Dea; Jurković, Slaven; Brambilla, Marco; Kalra, Mannudeep K.
Optimization of paranasal sinus CT procedure: Ultra-low dose CT as a roadmap for pre-functional endoscopic sinus surgery // Physica medica, 78 (2020), 195-200 doi:10.1016/j.ejmp.2020.09.014 (međunarodna recenzija, članak, znanstveni)
Diklić, A., Valković Zujić, P., Šegota, D., Dundara Debeljuh, D., Jurković, S., Brambilla, M. & Kalra, M. (2020) Optimization of paranasal sinus CT procedure: Ultra-low dose CT as a roadmap for pre-functional endoscopic sinus surgery. Physica medica, 78, 195-200 doi:10.1016/j.ejmp.2020.09.014.
@article{article, author = {Dikli\'{c}, Ana and Valkovi\'{c} Zuji\'{c}, Petra and \v{S}egota, Doris and Dundara Debeljuh, Dea and Jurkovi\'{c}, Slaven and Brambilla, Marco and Kalra, Mannudeep K.}, year = {2020}, pages = {195-200}, DOI = {10.1016/j.ejmp.2020.09.014}, keywords = {CT protocol, Paranasal sinuses, Radiation dose}, journal = {Physica medica}, doi = {10.1016/j.ejmp.2020.09.014}, volume = {78}, issn = {1120-1797}, title = {Optimization of paranasal sinus CT procedure: Ultra-low dose CT as a roadmap for pre-functional endoscopic sinus surgery}, keyword = {CT protocol, Paranasal sinuses, Radiation dose} }
@article{article, author = {Dikli\'{c}, Ana and Valkovi\'{c} Zuji\'{c}, Petra and \v{S}egota, Doris and Dundara Debeljuh, Dea and Jurkovi\'{c}, Slaven and Brambilla, Marco and Kalra, Mannudeep K.}, year = {2020}, pages = {195-200}, DOI = {10.1016/j.ejmp.2020.09.014}, keywords = {CT protocol, Paranasal sinuses, Radiation dose}, journal = {Physica medica}, doi = {10.1016/j.ejmp.2020.09.014}, volume = {78}, issn = {1120-1797}, title = {Optimization of paranasal sinus CT procedure: Ultra-low dose CT as a roadmap for pre-functional endoscopic sinus surgery}, keyword = {CT protocol, Paranasal sinuses, Radiation dose} }