Comparison of Two Quantitative Methods for Determination of Dabigatran Concentration: Diluted Thrombin Time Test and Innovance Dabigatran Test in Real Life Patients Treated with Dabigatran
Comparison of Two Quantitative Methods for Determination of Dabigatran Concentration: Diluted Thrombin Time Test and Innovance Dabigatran Test in Real Life Patients Treated with Dabigatran // Res Pract Thromb Haemost / Cushman, Mary (ur.).
Medford: John Wiley & Sons, 2020. PB0560, 1 (poster, međunarodna recenzija, sažetak, znanstveni)
CROSBI ID: 1073698 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Comparison of Two Quantitative Methods for
Determination of Dabigatran Concentration:
Diluted Thrombin Time Test and Innovance
Dabigatran Test in Real Life Patients Treated
with Dabigatran
Autori
Šupraha Goreta, Sandra ; Margetić, Sandra ; Ćelap, Ivana ; Mihić, Roman
Vrsta, podvrsta i kategorija rada
Sažeci sa skupova, sažetak, znanstveni
Izvornik
Res Pract Thromb Haemost
/ Cushman, Mary - Medford : John Wiley & Sons, 2020
Skup
Annual Meeting of the International Society on Thrombosis and Haemostasis (ISTH 2020)
Mjesto i datum
Online, 12.07.2020. - 14.07.2020
Vrsta sudjelovanja
Poster
Vrsta recenzije
Međunarodna recenzija
Ključne riječi
dabigatran, diluted thrombin time test
Sažetak
Background: Clinical application of dabigatran does not require routine therapeutic monitoring. However, there are special clinical situations in which measurement of dabigatran concentration in circulation should be performed using specific quantitative methods. Aims: The aim of this study was to compare two quantitative methods for dabigatran concentration i.e. commercial chromogenic Innovance Dabigatran assay (DTI) and in house optimized coagulometric diluted thrombin time (dTT) test. Methods: Dabigatran concentrations were determined in 38 patients using two methods on BCS XP coagulation analyzer (Siemens Healthineers, Germany): Innovance Dabigatran assay (Siemens Healthineers, Germany) with drug specific calibrators (Siemens Healthineers, Germany) and dTT test as an in house applied protocol with thrombin time reagent (BC Thrombin, Siemens Healthineers, Germany) calibrated with commercial dabigatran calibrators (Diagnostica Stago, France). Passing-Bablok regression and Bland-Altman methods were used for method comparison and Wilcoxon test was used to test differences between pairs of samples.The study was funded by the Croatian Science Foundation as part of the research project IP- 2016-06-8208. Results: Passing and Bablok regression analysis (y=-7.653 (-15.085 to -2.782) +0.986x (0.939 to1.033) revealed constant difference between two methods (Figure 1). Bland-Altman analysis showed statistically significant mean difference of 7.7% (P< 0.001) between dTT assay compared with DTI. Further, comparison of the results for dabigatran concentrations using DTI and dTT assays showed statistically significant difference (P=0.0002), but without clinically significant difference (Table 1). Conclusions: Although two evaluated methods for quantitative determination of dabigatran concentration showed significant constant difference, that difference is not clinically significant, so both can be used interchangeably. However, commercial DTI assay could be recommended as a method of choice for quantitative measurement of dabigatran concentration, since it has much wider measurement range without sample dilution (up to 500 ng/mL) compared to dTT assay (up to 256 ng/mL), uses chromogenic principle and time of testing is much shorter.
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
POVEZANOST RADA
Projekti:
HRZZ-IP-2016-06-8208 - Novi oralni antikoagulansi: povezanost koncentracije lijeka i antikoagulantnog učinka (LAB-NOAC) (Margetić, Sandra, HRZZ - 2016-06) ( CroRIS)
Ustanove:
Farmaceutsko-biokemijski fakultet, Zagreb,
KBC "Sestre Milosrdnice"
Citiraj ovu publikaciju:
Časopis indeksira:
- Web of Science Core Collection (WoSCC)
- Emerging Sources Citation Index (ESCI)