Pregled bibliografske jedinice broj: 1048547
Cochlear implant electrode array extrusion – is reimplantation feasible? – case report
Cochlear implant electrode array extrusion – is reimplantation feasible? – case report // Medicina Fluminensis : Medicina Fluminensis, 55 (2019), 1; 99-104 (domaća recenzija, prikaz, znanstveni)
CROSBI ID: 1048547 Za ispravke kontaktirajte CROSBI podršku putem web obrasca
Naslov
Cochlear implant electrode array extrusion – is
reimplantation feasible? – case report
Autori
Velepič, Marko ; Manestar, Dubravko ; Valković Zujić, Petra ; Malvić, Goran ; Zubović Velepič, Sanja
Izvornik
Medicina Fluminensis : Medicina Fluminensis (1847-6864) 55
(2019), 1;
99-104
Vrsta, podvrsta i kategorija rada
Radovi u časopisima, prikaz, znanstveni
Ključne riječi
chronic otitis media ; cochlear implant ; extracochlear electrode extrusion
Sažetak
Aim: To present two cases of late complications of cochlear implantation. Case report: In both cases extracochlear extrusion of the implant electrode occurred. More than ten years after successful cochlear implantation, the hearing performance gradually worsened to the point of no intelligibility in both patients. Otomicroscopy revealed chronic otitis media (COM) with an electrode in the external ear canal in implanted ears of both patients. Extracochlear electrode extrusion was noticed on CT scans in both cases. In the first case the device was explanted, the electrode was left in the cochlea and tympanoplasty was performed. Regardless the normal status of the middle ear, extracochlear extrusion occurred again. Conclusion: After extracochlear electrode extrusion, the reimplantation is likely to be unsuccessful.
Izvorni jezik
Engleski
Znanstvena područja
Kliničke medicinske znanosti
POVEZANOST RADA
Ustanove:
Medicinski fakultet, Rijeka,
Klinički bolnički centar Rijeka
Citiraj ovu publikaciju:
Časopis indeksira:
- Scopus