Analysis of Coupling between 900 MHz Mobile Telephone and Implanted Pacemaker (CROSBI ID 486672)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Hrabar, Silvio ; Balachandran, Wamadeva ; Bartolić, Juraj
engleski
Analysis of Coupling between 900 MHz Mobile Telephone and Implanted Pacemaker
The worst-case electromagnetic coupling between a 900 MHz mobile telephone and implanted pacemaker with unipolar catheter was investigated by computer simulation. The model comprises a dielectric block with embedded pacemaker and a half-wavelength dipole situated above the surface. A pacemaker connector was modelled as a small loop and a catheter as an insulated linear antenna. This model was combined with Finite Difference Time Domain calculation of the field induced within dielectric block. The analysis with 2W transmitter gave the peak interference voltage of 1 V for the connector and 0.3 V for the catheter. These results show that the pacemaker connector is more responsible for coupling than a catheter.
pacemaker; electromagnetic compatibility; interference; mobile telephone
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
365-368-x.
2001.
objavljeno
Podaci o matičnoj publikaciji
Proceeding on European Microwave Conference 2001
Nightingale, S. ; Philippakis, M.
London : Delhi: European Microwave Association
Podaci o skupu
European Microwave Conference
predavanje
25.09.2001-27.09.2001
London, Ujedinjeno Kraljevstvo