Development of a New Potentiometric Sensor based on home made Iodide ISE Enriched with ZnO Nanoparticles and its Application for Determination of Penicillamine (CROSBI ID 257904)
Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Prkić, Ante ; Vukušić, Tina ; Giljanović, Josipa ; Sokol, Vesna ; Bošković, Perica ; Lučić Lavčević, Magdy ; Mitar, Ivana ; Jakić, Miće
engleski
Development of a New Potentiometric Sensor based on home made Iodide ISE Enriched with ZnO Nanoparticles and its Application for Determination of Penicillamine
This paper describes the preparation and application of “home-made” iodide ISE enriched with ZnO nanoparticles. Iodide ISE membrane was made of AgI:Ag2S:PTFE = 1:1:2 and enriched with ZnO nanoparticles in ratio 1-5.0 wt.%. Prepared membranes were used for penicillamine (Pen) determination in acetic buffer (pH = 4-4.75) and in perchloric acid (pH = 1-2). Due to the low price method, simplicity and relative speed, a possibility for Pen determination was found. Pen determination showed better results in an acetic buffer, pH = 4 than in perchloric acid. All experiments for Pen measurements were done without pretreatment of pharmaceuticals. The determination is based on the reaction between Pen with both Ag+ and Zn2+ from the electrode membrane. A newly described method has linear response range for Pen 2.45×10−6-1×10−2 mol L−1 and a detection limit of 2.24×10−6 mol L−1. The found concentrations of Pen are in a very good agreement with the declared ones with the standard deviation of 2.70 %.
penicillamine, potentiometric, determination, “home-made” ion-selective electrode, nanoparticles
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o izdanju
13 (11)
2018.
10894-10903
objavljeno
1452-3981
10.20964/2018.11.92