Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures (CROSBI ID 247289)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

(EURADOS Working Group 12 Dosimetry in medical imaging) Järvinen, Hannu ; Farad, Jad ; Siiskonen, Teemu ; Ciraj-Bjelac, Olivera ; Dabin, Jeremie ; Carinou, Eleftheria ; Domienik, Joanna ; Kluszczynski, Dariusz ; Knežević, Željka ; Kopec, Renata et al. Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures // Physica medica, 46 (2018), 67-74. doi: 10.1016/j.ejmp.2018.01.010

Podaci o odgovornosti

Järvinen, Hannu ; Farad, Jad ; Siiskonen, Teemu ; Ciraj-Bjelac, Olivera ; Dabin, Jeremie ; Carinou, Eleftheria ; Domienik, Joanna ; Kluszczynski, Dariusz ; Knežević, Željka ; Kopec, Renata ; Majer, Marija ; Malchair, Francoise ; Negri, Anna ; Pankowski, Piotr ; Sarmento, Sandra ; Trianni, Annalisa

EURADOS Working Group 12 Dosimetry in medical imaging

engleski

Feasibility of setting up generic alert levels for maximum skin dose in fluoroscopically guided procedures

Purpose The feasibility of setting-up generic, hospital-independent dose alert levels to initiate vigilance on possible skin injuries in interventional procedures was studied for three high-dose procedures (chemoembolization (TACE) of the liver, neuro-embolization (NE) and percutaneous coronary intervention (PCI)) in 9 European countries. Methods Gafchromic® films and thermoluminescent dosimeters (TLD) were used to determine the Maximum Skin Dose (MSD). Correlation of the online dose indicators (fluoroscopy time, kerma- or dose-area product (KAP or DAP) and cumulative air kerma at interventional reference point (Ka, r)) with MSD was evaluated and used to establish the alert levels corresponding to a MSD of 2 Gy and 5 Gy. The uncertainties of alert levels in terms of DAP and K a, r, and uncertainty of MSD were calculated. Results About 20-30 % of all MSD values exceeded 2 Gy while only 2-6 % exceeded 5 Gy. The correlations suggest that both DAP and Ka, r can be used as a dose indicator for alert levels (Pearson correlation coefficient p mostly > 0.8), while fluoroscopy time is not suitable (p mostly <0.6). Generic alert levels based on DAP (Gy cm2) were suggested for MSD of both 2 Gy and 5 Gy (for 5 Gy: TACE 750, PCI 250 and NE 400). The suggested levels are close to the lowest values published in several other studies. The uncertainty of the MSD was estimated to be around 10 -15 % and of hospital-specific skin dose alert levels about 20-30 % (with coverage factor k=1). Conclusions The generic alert levels are feasible for some cases but should be used with caution, only as the first approximation, while hospital-specific alert levels are preferred as the final approach.

interventional radiology ; maximum skin dose ; online dose indicator ; skin dose alert

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

46

2018.

67-74

objavljeno

1120-1797

1724-191X

10.1016/j.ejmp.2018.01.010

Povezanost rada

Interdisciplinarne prirodne znanosti, Temeljne medicinske znanosti

Poveznice
Indeksiranost