Short –therm stability of routine haematology parameters after refrigeration.
BACKGROUND-AIM: Storage time and temperature may significantly affect routine haematology results. There is no unambiguous criteria for whole blood stability at 2-8°C. Our aim was to determine short-therm stability of routine haematology parameters in whole blood samples stored in different time intervals within 48 hours at 2-8 °C. METHODS: Blood was collected from 20 patients using K3EDTA Vacuette tubes (Greiner Bio-One GmbH, Kremsmünster, Austria). Routine haematology parameters were measured using the Sysmex XT 1800i analyzer (Sysmex Corporation, Kobe, Japan) immediately upon receipt. Whole blood samples where stored at 2-8°C and re- analysed after a 6, 12, 24 and 48 hours after baseline measurements. Paired samples t-test was used for statistical comparisons to baseline results. The level of significance was set at P<0.05. Mean biases from baseline results were compared to Ricos desirable biological specifications to define clinically relevant variations. RESULTS: Statistically different results compared to baseline values were found for: WBC after 6, 12 and 48 hours (P=0.008, P=0.035 and P=0.012, respectively) ; hematocrit after 48 hours (P=0.031) ; MCV after 24 and 48 hours (P=0.008 and P<0.001) ; MCHC after 48 hours (P<0.001) ; platelets after 6 hours (P=0.038) ; MPV after 6, 12, 24 and 48 hours (P<0.001) ; neutrophils (%), lymphocytes (%) and eosinophils (%) after 48 hours storage (P=0.001, P=0.016 and P=0.015, respectively). MPV exceeded the predefined criteria of 2.3% for clinical significance in all the time points tested (4.9% after 6 ; 7.5% after 12 ; 10.3% after 24 and 14.1% after 48 hours). WBC, MCHC and lymphocytes (%) exceeded the criteria for desirable bias after 24 hours storage. CONCLUSION: All the routine haematology parameters tested, except MPV, showed acceptable stability when stored at 2-8 °C within 24 hours, compared to predefined desirable specification for bias in our laboratory conditions.
temperature stability, hematologic parameters, hematologic analyser
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
37-38.
2017.
nije evidentirano
objavljeno
10.1515/cclm-2017-0137
Podaci o matičnoj publikaciji
Clinical chemistry and laboratory medicine
Berlin: Walter de Gruyter
1434-6621
1437-4331
Podaci o skupu
4th EFLM-BD European Conference on Preanalytical Phase, Amsterdam
poster
24.03.2017-25.03.2017
Amsterdam, Nizozemska
Povezanost rada
Farmacija, Kliničke medicinske znanosti