Kinetic Spectrophotometric Determination of N-Acetyl-L-cysteine Ethyl Ester (NACET) Generating Chromogenic Copper(I)Ln Complexes with Different Ligands


Kukoc-Modun, Lea; Tsikas, Dimitrios; Kraljević, Tomislav; Biocic, Maja; Radić, Njegomir
Kinetic Spectrophotometric Determination of N-Acetyl-L-cysteine Ethyl Ester (NACET) Generating Chromogenic Copper(I)Ln Complexes with Different Ligands // Croatica chemica acta, 90 (2017), 2; 263-271 doi:10.5562/cca3135 (međunarodna recenzija, članak, znanstveni)


Naslov
Kinetic Spectrophotometric Determination of N-Acetyl-L-cysteine Ethyl Ester (NACET) Generating Chromogenic Copper(I)Ln Complexes with Different Ligands

Autori
Kukoc-Modun, Lea ; Tsikas, Dimitrios ; Kraljević, Tomislav ; Biocic, Maja ; Radić, Njegomir

Izvornik
Croatica chemica acta (0011-1643) 90 (2017), 2; 263-271

Vrsta, podvrsta i kategorija rada
Radovi u časopisima, članak, znanstveni

Ključne riječi
N-acetyl-L-cysteine ethyl ester ; kinetic spectrophotometry ; neocuproine ; bicinchoninic acid ; bathocuproine disulfonic acid

Sažetak
Three simple, sensitive and robust kinetic spectrophotometric methods for the determination of a novel lipophilic thiol compound, N-acetyl-L-cysteine ethyl ester (NACET), have been developed and validated. The methods are based on the reduction of Cu(II)-ligand complex to Cu(I)-ligand complex with the analyte. Studied ligands were neocuproine, bicinchoninic acid (BCA) and bathocuproine disulfonic acid (BCS). The development of chromogenic complexes was followed using kinetic setup with spectrophotometric detection at 458, 562 and 483 nm for the reactions of NACET with neocuproine, BCA and BCS, respectively. The calculated reaction orders with respect to NACET concentration were found to be 1.07, 1.01, 1.07, respectively, thus confirming a first order of reaction. The initial rate and fixed time methods were utilized for constructing calibration curves. Assays limits of detection were 1.4 × 10–7, 3.2 × 10−7 and 6.0 × 10−8 mol L–1, respectively. The analytical performance of the methods, in terms of accuracy and precision, was established.

Izvorni jezik
Engleski

Znanstvena područja
Kemija



POVEZANOST RADA


Projekt / tema
011-0000000-3217 - Određivanje tiola primjenom potenciometrijskih senzora i spektroskopije (Njegomir Radić, )

Ustanove
Kemijsko-tehnološki fakultet, Split

Časopis indeksira:


  • Current Contents Connect (CCC)
  • Web of Science Core Collection (WoSCC)
    • SCI-EXP, SSCI i/ili A&HCI


Citati