Inspection and Testing of Diagnostic Ultrasound Devices (CROSBI ID 59686)

Prilog u knjizi | ostalo | međunarodna recenzija

Žauhar, Gordana ; Diklić, Ana ; Jurković, Slaven Inspection and Testing of Diagnostic Ultrasound Devices // Inspection of Medical Devices For Regulatory Purposes / Badnjević, Almir ; Cifrek, Mario ; Magjarević, Ratko et al. (ur.). Singapur: Springer, 2018. str. 115-140

Podaci o odgovornosti

Žauhar, Gordana ; Diklić, Ana ; Jurković, Slaven

engleski

Inspection and Testing of Diagnostic Ultrasound Devices

This chapter provides an overview of the safety aspects of application of ultrasound in medicine. It starts with the short history of ultrasound methods and devices as well as basic principles of ultrasound imaging systems. The application of ultrasound in medicine greatly evolved and nowadays it can be divided into two main areas: imaging and therapy. In order to assure a safe and responsible application of ultrasound in medicine one should be aware of physical processes which can be produced in tissue by ultrasound such as temperature rise, cavitation and acoustic streaming. The importance of understanding how these processes can affect the human cell is self-explanatory. In order to better understand the guidelines for testing and quality control of ultrasonic devices it is necessary to give an overview of basic output parameters. Only the most important parameters from the point of safe use of ultrasound are described, e.g. acoustic pressure, acoustic power and intensity. In order to protect the public against inappropriate exposure when ultrasound is used for medical applications, international standards and national regulations are developed. Diagnostic ultrasound imaging is very often the basis for diagnostic decision ; therefore it is also necessary to include such systems into a comprehensive quality assurance programme. Ultrasound systems used for therapy have larger intensities though there are additional safety requirements compared to diagnostic systems. The ultrasound intensity, effective radiation area and beam non-uniformity ratio and are parameters which should be monitored.

ultrasound devices, safety

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

115-140.

objavljeno

Podaci o knjizi

Inspection of Medical Devices For Regulatory Purposes

Badnjević, Almir ; Cifrek, Mario ; Magjarević, Ratko ; Džemić, Zijad

Singapur: Springer

2018.

978-981-10-6649-8

1864-5763

Povezanost rada

Fizika