Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi !

Experimental Method for Determination of Self- Heating at the Point of Measurement (CROSBI ID 243857)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Sestan, Danijel ; Zvizdic, Davor ; Grgec-Bermanec, Lovorka Experimental Method for Determination of Self- Heating at the Point of Measurement // International journal of thermophysics, 38 (2017), 9; 132, 11. doi: 10.1007/s10765-017-2268-9

Podaci o odgovornosti

Sestan, Danijel ; Zvizdic, Davor ; Grgec-Bermanec, Lovorka

engleski

Experimental Method for Determination of Self- Heating at the Point of Measurement

This paper presents a new experimental method and algorithm for the determination of self- heating of platinum resistance thermometer (PRT) when the temperature instability of medium of interest would prevent an accurate self-heating determination using standard methods. In temperature measurements performed by PRT, self- heating is one of the most common sources of error and arises from the increase in sensor temperature caused by the dissipation of electrical heat when measurement current is applied to the temperature sensing element. This increase depends mainly on the applied current and the thermal resistances between thermometer sensing element and the environment surrounding the thermometer. The method is used for determination of self-heating of a 100 ΩΩ industrial PRT which is intended for measurement of air temperature inside the saturation chamber of the primary dew/frost point generator at the Laboratory for Process Measurement (HMI/FSB-LPM). Self-heating is first determined for conditions present during the comparison calibration of the thermometer, using the calibration bath. The measurements were then repeated with thermometer being placed in an air stream inside the saturation chamber. The experiment covers the temperature range between −65∘C−65∘C and 10∘C10∘C . Self- heating is determined for two different air velocities and two different vertical positions of PRT in relation to the chamber bottom.

Platinum resistance thermometers ; Self-heating ; Uncertainty

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

38 (9)

2017.

132

11

objavljeno

0195-928X

10.1007/s10765-017-2268-9

Povezanost rada

Fizika, Temeljne tehničke znanosti

Poveznice
Indeksiranost