Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Comparison of response of passvie dosimetry systems in scanning proton radiotherpy - a study using paediatric anthropomorphic phantoms (CROSBI ID 243676)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

(EURADOS Working Group) Knežević, Željka ; Ambrozova, Iva ; Domingo, Carles ; De Saint-Hubert, Marijke ; Majer, Marija ; Martínez-Rovira, Imma ; Miljanić, Saveta ; Mojżeszek, Natalia ; Porwół, Paulina ; Ploc, Ondrej et al. Comparison of response of passvie dosimetry systems in scanning proton radiotherpy - a study using paediatric anthropomorphic phantoms // Radiation protection dosimetry, 180 (2018), 1-4; 256-260. doi: 10.1093/rpd/ncx254

Podaci o odgovornosti

Knežević, Željka ; Ambrozova, Iva ; Domingo, Carles ; De Saint-Hubert, Marijke ; Majer, Marija ; Martínez-Rovira, Imma ; Miljanić, Saveta ; Mojżeszek, Natalia ; Porwół, Paulina ; Ploc, Ondrej ; Romero-Expósito, Maite ; Stolarczyk, Liliana ; Trinkl, Sebastian ; Harrison, Roger M. ; Olko, Pawel

EURADOS Working Group

engleski

Comparison of response of passvie dosimetry systems in scanning proton radiotherpy - a study using paediatric anthropomorphic phantoms

Proton beam therapy has advantages in comparison to conventional photon radiotherapy due to the physical properties of proton beams (e.g. sharp distal fall off, adjustable range and modulation). In proton therapy, there is the possibility of sparing healthy tissue close to the target volume. This is especially important when tumors are located next to critical organs and while treating cancer in paediatric patients. On the other hand, the interactions of protons with matter result in the production of secondary radiation, mostly neutrons and gamma radiation, which deposit their energy at a distance from the target. The aim of this study was to compare the response of different passive dosimetry systems in mixed radiation field induced by proton pencil beam inside anthropomorphic phantoms representing 5 and 10 year-old children. Doses were measured in different organs with thermoluminescent (MTS-7, MTS-6 and MCP-N), radiophotoluminescent (GD-352M and GD302-M), bubble and poly-allyl-diglycol carbonate (PADC) track detectors. Results show that RPL detectors are the less sensitive for neutrons than LiF TLDs and can be applied for in-phantom dosimetry of gamma component. Neutron doses determined using track detectors, bubble detectors and pairs of MTS-7/MTS-6 are consistent within the uncertainty range. This is the first study dealing with measurements on child anthropomorphic phantoms irradiated by a pencil scanning beam technique.

scanning proton radiotherapy ; TLD ; RPL ; bubble and track detectors ; anthropomorphic child phantoms

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

180 (1-4)

2018.

256-260

objavljeno

0144-8420

1742-3406

10.1093/rpd/ncx254

Povezanost rada

Fizika, Kemija, Kliničke medicinske znanosti

Poveznice
Indeksiranost