Organ Doses from Pediatric Head CT Scan (CROSBI ID 653350)

Prilog sa skupa u časopisu | sažetak izlaganja sa skupa | međunarodna recenzija

Liu, H. ; Liu, Q. ; Majer, Marija ; Knežević, Željka ; Miljanić, Saveta ; Hršak, Hrvoje ; Zhuo, W. Organ Doses from Pediatric Head CT Scan // Medical physics. 2016. str. 3394-3394 doi: 10.1118/1.4955860

Podaci o odgovornosti

Liu, H. ; Liu, Q. ; Majer, Marija ; Knežević, Željka ; Miljanić, Saveta ; Hršak, Hrvoje ; Zhuo, W.

engleski

Organ Doses from Pediatric Head CT Scan

Purpose: To evaluate the organ doses of pediatric patients who undergoing head CT scan using Monte Carlo (MC) simulation and compare it with measurements in anthropomorphic child phantom.. Methods: A ten years old children voxel phantom was developed from CT images, the voxel size of the phantom was 2mm*2mm*2mm. Organ doses from head CT scan were simulated using MCNPX software, 180 detectors were placed in the voxel phantom to tally the doses of the represented tissues or organs. When performing the simulation, 120 kVp and 88 mA were selected as the scan parameters. The scan range covered from the top of the head to the end of the chain, this protocol was used at CT simulator for radiotherapy. To validate the simulated results, organ doses were measured with radiophotoluminescence (RPL) detectors, placed in the 28 organs of the 10 years old CIRS ATOM phantom. Results: The organ doses results matched well between MC simulation and phantom measurements. The eyes dose was showed to be as expected the highest organ dose: 28.11 mGy by simulation and 27.34 mGy by measurement respectively. Doses for organs not included in the scan volume were much lower than those included in the scan volume, thymus doses were observed more than 10 mGy due the CT protocol for radiotherapy covered more body part than routine head CT scan. Conclusion: As the eyes are superficial organs, they may receive the highest radiation dose during the CT scan. Considering the relatively high radio sensitivity, using shielding material or organ based tube current modulation technique should be encouraged to reduce the eye radiation risks. Scan range was one of the most important factors that affects the organ doses during the CT scan. Use as short as reasonably possible scan range should be helpful to reduce the patient radiation dose. This work was supported by the National Natural Science Foundation of China(11475047)

Dosimetry ; Computed tomography ; Monte Carlo methods ; Eyes ; Image sensors ; Tissues ; Medical imaging

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o prilogu

3394-3394.

2016.

nije evidentirano

objavljeno

10.1118/1.4955860

Podaci o matičnoj publikaciji

Medical physics

0094-2405

2473-4209

Podaci o skupu

Fifty-eighth annual meeting of the american association of physicists in medicine

poster

31.07.2016-04.08.2016

Washington D.C., Sjedinjene Američke Države

Povezanost rada

Fizika, Kliničke medicinske znanosti

Poveznice
Indeksiranost