The effect of counting conditions on pure beta emitter determination by Cherenkov counting (CROSBI ID 230151)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Coha, Ivana ; Neufuss, Soběslav ; Grahek, Željko ; Mojmír ; Nodilo, Marijana ; John, Jan The effect of counting conditions on pure beta emitter determination by Cherenkov counting // Journal of radioanalytical and nuclear chemistry, 310 (2016), 2; 891-903. doi: 10.1007/s10967-016-4853-y

Podaci o odgovornosti

Coha, Ivana ; Neufuss, Soběslav ; Grahek, Željko ; Mojmír ; Nodilo, Marijana ; John, Jan

engleski

The effect of counting conditions on pure beta emitter determination by Cherenkov counting

The Cherenkov counting efficiencies measured by three commercially available liquid scintillation counters, Triathler, Tri-Carb 3180, Hidex 300 SL were compared. The influences of energy of pure beta emitters, volume, and solvent as well as instrument type on Cherenkov counting efficiency were studied. The best Cherenkov counting efficiencies are on the Hidex 300 SL and when methanol is used as a solvent. It could be concluded that the sensitivities of low-level versions of Tri-Carb and Hidex 300 SL are comparable. Based on the overall results obtained it has been concluded that the TDCR Cherenkov counting is well suited for routine quantitative determination of low activities of high energetic beta emitters.

TDCR Cherenkov efficiency ; Pure beta emitters ; Hidex 300 SL ; Tri-Carb 3180 TR/SL ; Triathler multilabel counter

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

310 (2)

2016.

891-903

objavljeno

0236-5731

10.1007/s10967-016-4853-y

Povezanost rada

Fizika, Kemija

Poveznice
Indeksiranost