Multi-entrance fixed point filled with aluminium (CROSBI ID 637346)
Prilog sa skupa u zborniku | sažetak izlaganja sa skupa | međunarodna recenzija
Podaci o odgovornosti
Zvizdić, Davor ; Šestan, Danijel ; Šariri, Kristina
engleski
Multi-entrance fixed point filled with aluminium
This paper presents the results of testing of the multi-entrance fixed point cell (MEFP) filled with aluminium. The MEFP cells aim to improve the performances of the zones of controlled temperature used for calibration of thermometers by comparison method. The improvements are achieved by using the phase transition of metals to reduce the uncertainties due to temporal instability and gradients in those zones. Since the MEFP itself does not represent the temperauture standard, the metals of modest purity are acceptable for its manufacture.
thermometry; fixed point; comparison calibration
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
165-x.
2016.
objavljeno
Podaci o matičnoj publikaciji
TEMPMEKO 2016 - Book of Abstracts
Aleksandra Kowal
Wrocław: Instytut niskich temperatur i badán strukturalnych, Wroclaw
978-83-939559-7-8
Podaci o skupu
TEMPMEKO 2016 - XIII international Symposium on Temperature and Thermal Measurement in Industry and Science
poster
26.06.2016-01.07.2016
Zakopane, Poljska