A comparison of height-based and slope-based criteria to determine arterial pressure in finger (CROSBI ID 482802)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Lacković, Igor
engleski
A comparison of height-based and slope-based criteria to determine arterial pressure in finger
Two types of criteria used to determine systolic and diastolic pressure in finger by oscillometric method were compared in terms of accuracy and reliability. Experimental evaluation was performed on 32 volunteers with manual auscultatory method used as a reference. The results may be summarised as follows: for height-based criteria a good agreement with reference measurements was found for both systolic and diastolic pressure ; for slope-based criteria a good agreement was found only for systolic pressure, whereas this approach tends to overestimate diastolic pressure by more than 10 mmHg.
oscillometric method; systolic pressure; diastolic pressure; finger blood pressure
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
50-51.
2002.
objavljeno
Podaci o matičnoj publikaciji
Sigurdsson, Stefan, B.
Reykjavík: Icelandic Society of Biomedical Engineering and Medical Physics
Podaci o skupu
Nordic Baltic Conference on Biomedical Engineering and Medical Physics(12 ; 2002)
poster
18.06.2002-22.06.2002
Reykjavík, Island