Nalazite se na CroRIS probnoj okolini. Ovdje evidentirani podaci neće biti pohranjeni u Informacijskom sustavu znanosti RH. Ako je ovo greška, CroRIS produkcijskoj okolini moguće je pristupi putem poveznice www.croris.hr
izvor podataka: crosbi

Short-term tests of Potassium dIdD vector magnetometer (CROSBI ID 215106)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Vujić, Eugen Short-term tests of Potassium dIdD vector magnetometer // Acta geophysica, 63 (2015), 5; 1276-1295. doi: 10.1515/acgeo-2015-0039

Podaci o odgovornosti

Vujić, Eugen

engleski

Short-term tests of Potassium dIdD vector magnetometer

In this paper the results of short-term tests of a suspended Potassium dIdD vector magnetometer are presented. The instrument was installed on a stable pillar during May 2012 near Milton (Ontario, Canada). The aim was to investigate primarily the bias current stability and the noise level of the instrument (for measured components less than 10 pT in a rootmean- square sense). A brief description of the measurement principles and a review of formulae used for field component calculations are given. The errors of components, which arise from assumptions on dIdD orientation as well as from the total field time variations during a measurement cycle, are estimated and discussed, both for real and simulated data. Finally the accuracy of the instrument was estimated, and the maximal theoretical accuracies of components were better than 60 pT at mid-latitudes.

Potassium dIdD vector magnetometer ; magnetometer calibration ; error calculation

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

63 (5)

2015.

1276-1295

objavljeno

1895-6572

1895-7455

10.1515/acgeo-2015-0039

Povezanost rada

Geologija

Poveznice
Indeksiranost