Calibration of the Standard Resistors by means of the Standard Capacitors and Precise Digital Voltmeters (CROSBI ID 464994)
Prilog sa skupa u zborniku | izvorni znanstveni rad | međunarodna recenzija
Podaci o odgovornosti
Bego, Vojislav ; Butorac, Josip ; Hrsan, Davor
engleski
Calibration of the Standard Resistors by means of the Standard Capacitors and Precise Digital Voltmeters
The calibration is carried out by measuring the ratios of voltages on an auxiliary resistor, once as a result of the current flowing through the previously calibrated capacitance standard, and the other time as a result of the current through the resistance standard. Using these voltage ratios, obtained with DC and low frequency AC current, measured by the precise digital voltmeters, a resistance standard can be calibrated with the uncertainty of 0,2 ppm.
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
nije evidentirano
Podaci o prilogu
392-393-x.
1996.
objavljeno
Podaci o matičnoj publikaciji
CPEM '96 Conference Digest
Braun, Andreas
IEEE, Inc.
Podaci o skupu
Conference on Precision Electromagnetic Measurements
poster
17.06.1996-20.06.1996
Braunschweig, Njemačka