Correction of measured Gamma-Knife output factors for angular dependence of diode detectors and PinPoint ionization chamber (CROSBI ID 210157)

Prilog u časopisu | izvorni znanstveni rad | međunarodna recenzija

Hršak, Hrvoje ; Majer, Marija ; Grego, Timor ; Bibić, Juraj ; Heinrich, Zdravko Correction of measured Gamma-Knife output factors for angular dependence of diode detectors and PinPoint ionization chamber // Physica medica, 30 (2014), 8; 914-919. doi: 10.1016/j.ejmp.2014.09.002

Podaci o odgovornosti

Hršak, Hrvoje ; Majer, Marija ; Grego, Timor ; Bibić, Juraj ; Heinrich, Zdravko

engleski

Correction of measured Gamma-Knife output factors for angular dependence of diode detectors and PinPoint ionization chamber

Dosimetry for Gamma-Knife requires detectors with high spatial resolution and minimal angular dependence of response. Angular dependence and end effect time for p-type silicon detectors (PTW Diode P and Diode E) and PTW PinPoint ionization chamber were measured with Gamma- Knife beams. Weighted angular dependence correction factors were calculated for each detector. The Gamma- Knife output factors were corrected for angular dependence and end effect time. For Gamma-Knife beams angle range of 84°-54°. Diode P shows considerable angular dependence of 9% and 8% for the 18 mm and 14, 8, 4 mm collimator, respectively. For Diode E this dependence is about 4% for all collimators. PinPoint ionization chamber shows angular dependence of less than 3% for 18, 14 and 8 mm helmet and 10% for 4 mm collimator due to volumetric averaging effect in a small photon beam. Corrected output factors for 14 mm helmet are in very good agreement (within ±0.3%) with published data and values recommended by vendor (Elekta AB, Stockholm, Sweden). For the 8 mm collimator diodes are still in good agreement with recommended values (within ±0.6%), while PinPoint gives 3% less value. For the 4 mm helmet Diodes P and E show over-response of 2.8% and 1.8%, respectively. For PinPoint chamber output factor of 4 mm collimator is 25% lower than Elekta value which is generally not consequence of angular dependence, but of volumetric averaging effect and lack of lateral electronic equilibrium. Diodes P and E represent good choice for Gamma-Knife dosimetry.

Gamma-Knife output factor; Silicon diode; PinPoint ionization chamber; Correction for angular dependence

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

nije evidentirano

Podaci o izdanju

30 (8)

2014.

914-919

objavljeno

1120-1797

10.1016/j.ejmp.2014.09.002

Povezanost rada

Fizika, Kliničke medicinske znanosti