Solution to directivity problems in ultrasonically marked cardiac catheters (CROSBI ID 93014)

Prilog u časopisu | izvorni znanstveni rad

Breyer, Branko ; Ferek-Petrić, Božidar Solution to directivity problems in ultrasonically marked cardiac catheters // Automatika : časopis za automatiku, mjerenje, elektroniku, računarstvo i komunikacije, 40 (1999), 1-2; 7-9-x

Podaci o odgovornosti

Breyer, Branko ; Ferek-Petrić, Božidar

engleski

Solution to directivity problems in ultrasonically marked cardiac catheters

The problem of techological compromise between sensitivity and desirable broad directivity characteristic of ultrasonically marked catheter has been solved by application of divergent lenses to transmitting-receiving marker transducer. It was shown that the lenses can be designed in various ways by including asymmetric lenses that modify the elevation angle of transucer directivity.

catheter; guidance; procedure guidance; minimum invasive surgery; transducer directivity; ultrasound imaging; ultrasonically marked catheter